Massive aspiration past the tracheal tube cuff caused by closed tracheal suction system
- PMID: 21300670
- DOI: 10.1177/0885066610392516
Massive aspiration past the tracheal tube cuff caused by closed tracheal suction system
Abstract
Background: Aspiration past the tracheal tube cuff has been recognized to be a risk factor for the development of ventilator-associated pneumonia (VAP). This study investigated the effect of closed tracheal suctioning on aspiration of fluid past the tracheal tube cuff in an in vitro benchtop model.
Methods: High-volume low pressure tube cuffs of 7.5 mm internal diameter (ID) were placed in a 22 mm ID artificial trachea connected to a test lung. Positive pressure ventilation (PPV) with 15 cm H₂O peak inspiratory pressure and 5 cm H₂O positive end-expiratory pressure (PEEP) was used. A closed tracheal suction system (CTSS) catheter (size 14Fr) was attached to the tracheal tube and suction was performed for 5, 10, 15, or 20 seconds under 200 or 300 cm H₂O suction pressures. Amount of fluid (mL) aspirated along the tube cuff and the airway pressure changes were recorded for each suction procedure. Fluid aspiration during different suction conditions was compared using Kruskal-Wallis and Mann-Whitney test (Bonferroni correction [α = .01]).
Results: During 10, 15, and 20 seconds suction, airway pressure consistently dropped down to -8 to -13 cm H₂O (P < .001) from the preset level. Fluid aspiration was never observed under PPV + PEEP but occurred always during suctioning. Aspiration along the tube cuff was higher with -300 cm H₂O than with -200 cm H₂O suction pressure (P < .001) and was much more during 15 and 20 seconds suction time as compared to 5 seconds (P < .001).
Conclusion: Massive aspiration of fluid occurs along the tracheal tube cuff during suction with the closed tracheal suction system.
Keywords: aspiration; closed tracheal suction system; endotracheal tube.
© SAGE Publications 2011.
Similar articles
-
Closed tracheal suction and fluid aspiration past the tracheal tube. Impact of tube cuff and airway pressure.Minerva Anestesiol. 2011 Feb;77(2):166-71. Epub 2011 Feb 1. Minerva Anestesiol. 2011. PMID: 21283067
-
Effect of Lanz pressure regulating valve on self-sealing mechanism and air leakage across the tracheal tube cuffs in a benchtop model.J Intensive Care Med. 2013 Jul-Aug;28(4):247-51. doi: 10.1177/0885066612452847. Epub 2012 Jul 24. J Intensive Care Med. 2013. PMID: 22833049
-
Effect of applying positive pressure with or without endotracheal suctioning during extubation: a laboratory study.Respir Care. 2014 Dec;59(12):1905-11. doi: 10.4187/respcare.03121. Epub 2014 Nov 25. Respir Care. 2014. PMID: 25425709
-
The LoTrach system: its role in the prevention of ventilator-associated pneumonia.Nurs Crit Care. 2008 Sep-Oct;13(5):260-8. doi: 10.1111/j.1478-5153.2008.00286.x. Nurs Crit Care. 2008. PMID: 18816312 Review.
-
New issues and controversies in the prevention of ventilator-associated pneumonia.Am J Respir Crit Care Med. 2010 Oct 1;182(7):870-6. doi: 10.1164/rccm.201001-0081CI. Epub 2010 May 6. Am J Respir Crit Care Med. 2010. PMID: 20448095 Review.
Cited by
-
Closed Endotracheal Suction Systems for COVID-19: Rapid Review.Interact J Med Res. 2023 Jan 10;12:e42549. doi: 10.2196/42549. Interact J Med Res. 2023. PMID: 36548950 Free PMC article. Review.
-
Endotracheal tube cuff--small important part of a big issue.J Clin Monit Comput. 2012 Feb;26(1):53-60. doi: 10.1007/s10877-011-9333-x. Epub 2012 Jan 3. J Clin Monit Comput. 2012. PMID: 22212414
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous