Radiation protection for an intra-operative X-ray device
- PMID: 21304003
- PMCID: PMC3473704
- DOI: 10.1259/bjr/29466902
Radiation protection for an intra-operative X-ray device
Abstract
Objectives: Therapeutic partial breast irradiation can be delivered intra-operatively using the Intrabeam 50 kVp compact X-ray device. Spherical applicators are added to the source to give an isotropic radiation dose. The low energy of this unit leads to rapid attenuation with distance, but dose rates are much greater than for diagnostic procedures.
Methods: To investigate the shielding requirements for this unit, attenuation measurements were carried out with manufacturer-provided tungsten-rubber sheets, lead, plasterboard and bricks. A prospective environmental dose rate survey was also conducted in the designated theatre.
Results: As a result of isotropic geometry, the scattered dose around shielding can be 1% of primary and thus often dominates measured dose rates compared with transmission. The absorbed dose rate of the unshielded source at 1 m was 11.6 mGy h(-1) but this was reduced by 95% with the shielding sheets. Measured values for the common shielding materials were similar to reference data for the attenuation of a 50 kVp diagnostic X-ray beam. Two lead screens were constructed to shield operators remaining in the theatre and an air vent into a service corridor. A lead apron would also provide suitable attenuation, although a screen allows greater flexibility for treatment operators. With these measures, staff doses were reduced to negligible quantities. Survey measurements taken during patient treatments confirmed no additional measures were required, but the theatre should be a controlled area and access restricted.
Conclusion: Results from this study and reference data can be used for planning other facilities.
Figures



Similar articles
-
Absorbed radiation dose of the female breast during diagnostic multidetector chest CT and dose reduction with a tungsten-antimony composite breast shield: preliminary results.Clin Radiol. 2008 Mar;63(3):278-88. doi: 10.1016/j.crad.2007.07.029. Epub 2007 Dec 4. Clin Radiol. 2008. PMID: 18275868
-
Radiation survey around a Liac mobile electron linear accelerator for intraoperative radiation therapy.J Appl Clin Med Phys. 2009 Apr 28;10(2):131-138. doi: 10.1120/jacmp.v10i2.2950. J Appl Clin Med Phys. 2009. PMID: 19458597 Free PMC article.
-
Dosimetry measurements with an intra-operative x-ray device.Phys Med Biol. 2010 Jun 21;55(12):N359-69. doi: 10.1088/0031-9155/55/12/N02. Epub 2010 May 26. Phys Med Biol. 2010. PMID: 20505225
-
Dosimetric characteristics of a low-kV intra-operative x-ray source: implications for use in a clinical trial for treatment of low-risk breast cancer.Med Phys. 2003 Sep;30(9):2424-31. doi: 10.1118/1.1595611. Med Phys. 2003. PMID: 14528964
-
Lead-rubber shielding effect on radiation dose to the gonads from a bilateral hand X-ray examination.Radiography (Lond). 2022 May;28(2):360-365. doi: 10.1016/j.radi.2021.12.013. Epub 2022 Jan 6. Radiography (Lond). 2022. PMID: 35000868
Cited by
-
INTRABEAM intraoperative radiotherapy combined with portal vein infusion chemotherapy for treating hepatocellular carcinoma with portal vein tumor thrombus.BMC Surg. 2020 Aug 1;20(1):174. doi: 10.1186/s12893-020-00836-2. BMC Surg. 2020. PMID: 32738882 Free PMC article.
-
Bulgarian Experience in Vaginal Electronic Brachytherapy for Gynecologic Cancers' Treatment-First Results.J Clin Med. 2024 Dec 23;13(24):7849. doi: 10.3390/jcm13247849. J Clin Med. 2024. PMID: 39768772 Free PMC article.
-
Intraoperative radiotherapy in early stage breast cancer: potential indications and evidence to date.Br J Radiol. 2015 May;88(1049):20140686. doi: 10.1259/bjr.20140686. Epub 2015 Mar 3. Br J Radiol. 2015. PMID: 25734489 Free PMC article. Review.
-
Intraoperative radiation therapy with 50 kV x-rays: A multi-institutional review.J Appl Clin Med Phys. 2024 Mar;25(3):e14272. doi: 10.1002/acm2.14272. Epub 2024 Jan 26. J Appl Clin Med Phys. 2024. PMID: 38279520 Free PMC article. Review.
-
Is It Safe to Use a Lead Screen During Hip Arthroscopy?Arthrosc Sports Med Rehabil. 2021 Feb 3;3(2):e329-e334. doi: 10.1016/j.asmr.2020.09.022. eCollection 2021 Apr. Arthrosc Sports Med Rehabil. 2021. PMID: 34027439 Free PMC article.
References
-
- Williams JR, Thwaites DI. Radiotherapy physics in practice. Oxford: Oxford University Press, 2000
-
- Beddar AS, Biggs PJ, Chang S, Ezzell GA, Faddegon BA, Hensley FW, et al. Intraoperative radiation therapy using mobile electron linear accelerators: Report of AAPM Radiation Therapy Committee Task Group No. 72. Med Phys 2006;33:1476–89 - PubMed
-
- Soriani A, Felici G, Fantini M, Paolucci M, Borla O, Evangelisti G, et al. Radiation protection measurements around a 12 MeV mobile dedicated IORT accelerator. Med Phys 2010;37:995–1003 - PubMed
-
- Vaidya JS, Baum M, Tobias JS, Morgan S, D'Souza D. The novel technique of delivering targeted intraoperative radiotherapy (Targit) for early breast cancer. Eur J Surg Oncol 2002;28:447–54 - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials