Comparative Study
. 1990 Feb;4(1):30-4.
doi: 10.1016/0888-6296(90)90443-j.

A clinical evaluation of pulse oximetry during thoracic surgery

Affiliations
Comparative Study

A clinical evaluation of pulse oximetry during thoracic surgery

D P Desiderio et al. J Cardiothorac Anesth. 1990 Feb.

Abstract

To evaluate the utility of pulse oximetry for monitoring oxygenation during thoracic surgery, pulse oximeter oxygen saturation (SpO2) values from the Nellcor N-100 (Nellcor Inc, Haywood, CA) and Novametrix model 500 (Medical Systems Inc, Wallingford, CT) were compared with simultaneous arterial saturation values (SaO2) in 20 patients. A total of 255 matched observations were recorded, and the data were divided for statistical analysis into preinduction of anesthesia and postinduction groups. The preinduction group showed a good correlation between SpO2 and SaO2 values, with both pulse oximeters consistently overestimating the SaO2. However, once anesthesia was induced, there was no longer any correlation for either of the pulse oximeters versus simultaneous SaO2 values, although on average, the SpO2 values were significantly higher than the corresponding SaO2 values. It was concluded that pulse oximetry is useful in following trends of oxygenation in patients with preexisting lung pathology undergoing thoracic surgery, but it cannot replace arterial blood gas sampling for the intraoperative management of respiratory function.

PubMed Disclaimer

Publication types