Newborn screening for isovaleric acidemia using tandem mass spectrometry: data from 1.6 million newborns

Affiliations
Free article

Newborn screening for isovaleric acidemia using tandem mass spectrometry: data from 1.6 million newborns

Regina Ensenauer et al. Clin Chem. 2011 Apr.
Free article

Abstract

Background: Electrospray ionization-tandem mass spectrometry (ESI-MS/MS) has been used in the Bavarian newborn screening (NBS) program since 1999. The use of ESI-MS/MS has led to the inclusion of isovaleric acidemia (IVA) into NBS. We retrospectively evaluated data on more than 1.6 million newborns screened during 9.5 years.

Methods: Acylcarnitines from whole blood spotted on filter paper were converted to their corresponding butyl esters, and the samples were analyzed by use of ESI-MS/MS with stable isotope labeled internal standards.

Results: A total of 24 individuals with IVA were detected by use of a multiparametric threshold criteria panel including isovalerylcarnitine (C5) and the ratios of C5 to octanoyl-, butyryl-, and propionylcarnitine. A cutoff set at the 99.99th percentile for isolated C5 or at the 99th percentile for C5 plus at least 2 ratios resulted in a positive predictive value for IVA screening of 7.0% and an overall recall rate of 0.024%. Adjusted reference ranges for age and birth weight were applied, and the incidence of IVA in the study population was calculated to be 1 in 67,000. Missed cases were not brought to our attention. IVA was also detectable in cord blood and early postnatal blood samples.

Conclusions: IVA can be reliably detected in NBS through acylcarnitine analysis in dried blood spots by using multiparametric threshold criteria. Further improvement (positive predictive value 13.0%, recall rate 0.01%) can be achieved by using more stringent recall criteria. In view of the potentially life-threatening natural course of IVA in early life, presymptomatic diagnosis may thus prevent mortality and morbidity.