In vitro assessment of heart valve bioprostheses by cardiovascular magnetic resonance: four-dimensional mapping of flow patterns and orifice area planimetry
- PMID: 21342775
- DOI: 10.1016/j.ejcts.2010.12.040
In vitro assessment of heart valve bioprostheses by cardiovascular magnetic resonance: four-dimensional mapping of flow patterns and orifice area planimetry
Abstract
Objective: The hemodynamics in proximity to stented aortic bioprostheses still differ from that under physiological conditions. This may prevent desired cardiac remodeling and promote aortic diseases. Further improvements in prosthetic technology require an accurate survey of the flow conditions on the prosthetic level and in the ascending aorta. Cardiovascular magnetic resonance (CMR) may have the potential to provide more information by determining the prosthetic orifice area and visualizing the intravascular flow dynamics. We tested the feasibility to better characterize the hemodynamics of various stented bioprostheses in a pulsatile flow phantom by using CMR.
Methods: The custom-made model consisting of a commercially available pump generating pulsatile flow, a tube system filled with a glycerin-water mixture, and a handcrafted bulbar-shaped cylinder holding the bioprostheses and simulating the aortic root, was located in a clinical 1.5T CMR system. In this study, 10 stented aortic bioprostheses were investigated (Perimount® 21, 23; Mitroflow® 19, 25; Hancock® 21, 23, 25; Mosaic® 21; Epic Supra® 21, 23). The prosthetic orifice area was visualized using steady-state free-precession cine imaging (spatial/temporal resolution 1.3×1.3×5 mm³/29 ms), quantified by manual planimetry and compared with published transthoracic echocardiographic data. Time-resolved three-dimensional phase-contrast flow mapping (1.8×1.8×3 mm³/45 ms) was applied to analyze the transprosthetic flow pattern.
Results: Visualization of the prosthetic orifice area and the transprosthetic flow pattern was feasible in all prostheses. All orifice areas obtained by CMR in vitro were within one standard deviation of the mean of the published reference values obtained by echocardiography in vivo. Turbulent flow with vortex formation occurred both in proximity to the prosthesis and on the 'ascending aortic' level. Larger prosthetic sizes led to decreased flow velocities, but not mandatorily to less turbulences.
Conclusions: CMR allowed for a detailed interrogation of the fluid dynamics of various heart valve bioprostheses in a pulsatile flow model. It is an attractive tool to define proprietary reference values of the orifice area under standardized conditions and provides novel information regarding the flow pattern in proximity to the prosthesis.
Copyright © 2011 European Association for Cardio-Thoracic Surgery. Published by Elsevier B.V. All rights reserved.
Similar articles
-
Feasibility of cardiovascular magnetic resonance to assess the orifice area of aortic bioprostheses.Circ Cardiovasc Imaging. 2009 Sep;2(5):397-404, 2 p following 404. doi: 10.1161/CIRCIMAGING.108.840967. Epub 2009 May 22. Circ Cardiovasc Imaging. 2009. PMID: 19808628
-
Assessment of mitral bioprostheses using cardiovascular magnetic resonance.J Cardiovasc Magn Reson. 2010 Jun 23;12(1):36. doi: 10.1186/1532-429X-12-36. J Cardiovasc Magn Reson. 2010. PMID: 20573227 Free PMC article.
-
Functional Assessment of Bioprosthetic Aortic Valves by CMR.JACC Cardiovasc Imaging. 2016 Jul;9(7):785-793. doi: 10.1016/j.jcmg.2015.08.025. Epub 2016 May 12. JACC Cardiovasc Imaging. 2016. PMID: 27184505
-
Clinical and hemodynamic implications of supra-annular implant of biological aortic valves.J Cardiovasc Med (Hagerstown). 2006 Jul;7(7):524-32. doi: 10.2459/01.JCM.0000234771.96324.28. J Cardiovasc Med (Hagerstown). 2006. PMID: 16801814 Review.
-
Which available transapical transcatheter valve fits into degenerated aortic bioprostheses?Interact Cardiovasc Thorac Surg. 2010 Jul;11(1):83-5. doi: 10.1510/icvts.2010.234625. Epub 2010 Apr 15. Interact Cardiovasc Thorac Surg. 2010. PMID: 20395247 Review.
Cited by
-
Real-time phase contrast magnetic resonance imaging for assessment of haemodynamics: from phantom to patients.Eur Radiol. 2016 Apr;26(4):986-96. doi: 10.1007/s00330-015-3897-7. Epub 2015 Jul 19. Eur Radiol. 2016. PMID: 26188655
-
Automatic segmentation and tracking of biological prosthetic heart valves.J Med Imaging (Bellingham). 2021 Jan;8(1):015501. doi: 10.1117/1.JMI.8.1.015501. Epub 2021 Feb 12. J Med Imaging (Bellingham). 2021. PMID: 33604410 Free PMC article.
-
Hypertrophic obstructive cardiomyopathy-left ventricular outflow tract shapes and their hemodynamic influences applying CMR.Int J Cardiovasc Imaging. 2024 Nov;40(11):2371-2381. doi: 10.1007/s10554-024-03242-4. Epub 2024 Sep 20. Int J Cardiovasc Imaging. 2024. PMID: 39302632 Free PMC article.
-
Significant intra-valvular pressure loss across EPIC SUPRA and perimount magna supra-annular designed aortic bioprostheses in patients with normal aortic size.Indian Heart J. 2017 Jan-Feb;69(1):87-92. doi: 10.1016/j.ihj.2016.06.001. Epub 2016 Jun 11. Indian Heart J. 2017. PMID: 28228313 Free PMC article.
-
In vitro evaluation of flow patterns and turbulent kinetic energy in trans-catheter aortic valve prostheses.MAGMA. 2018 Feb;31(1):165-172. doi: 10.1007/s10334-017-0651-y. Epub 2017 Sep 18. MAGMA. 2018. PMID: 28921052
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources