Treponema-specific tests for serodiagnosis of syphilis: comparative evaluation of seven assays

Affiliations
Comparative Study

Treponema-specific tests for serodiagnosis of syphilis: comparative evaluation of seven assays

M J Binnicker et al. J Clin Microbiol. 2011 Apr.

Abstract

The diagnosis of syphilis is challenging and often relies on serologic tests to detect treponemal or nontreponemal antibodies. Recently, the Centers for Disease Control and Prevention and the Association of Public Health Laboratories proposed an update to the syphilis serology testing algorithm, in which serum samples are first tested using a treponema-specific test and positive samples are analyzed with a nontreponemal assay. The goal of this study was to compare the performance of seven treponemal assays (BioPlex 2200 syphilis IgG [Bio-Rad, Hercules, CA], fluorescent treponemal antibody [FTA] assay [Zeus Scientific, Raritan, NJ], Treponema pallidum particle agglutination [TP-PA; Fujirebio Diagnostics, Malvern, PA], Trep-Sure enzyme immunoassay [EIA; Phoenix Biotech, Oakville, Ontario, Canada], Trep-Chek EIA [Phoenix Biotech], Trep-ID EIA [Phoenix Biotech], and Treponema ViraBlot IgG [Viramed Biotech AG, Planegg, Germany]) using serum samples (n = 303) submitted to our reference laboratory. In addition to testing with these 7 assays, all samples were tested by a rapid plasma reagin (RPR) assay and a treponemal IgM Western blot assay (Viramed ViraBlot). Compared to the FTA assay as the gold standard, the evaluated treponemal tests demonstrated comparable levels of performance, with percent agreement ranging from 95.4% (95% confidence interval, 92.3 to 97.3) for the Trep-Sure EIA to 98.4% (96.1 to 99.4) for the Trep-ID EIA. Compared to a "consensus of the test panel" (defined as at least 4 of 7 treponemal tests being in agreement), the percent agreement ranged from 95.7% (92.7 to 97.5) for Trep-Sure to 99.3% (97.5 to 99.9) for Trep-ID. These data may assist clinical laboratories that are considering implementing a treponemal test for screening or confirmatory purposes.

PubMed Disclaimer

References

    1. Association of Public Health Laboratories 2009. Laboratory diagnostic testing for Treponema pallidum—expert consultation meeting summary report, January 13-15, 2009. Association of Public Health Laboratories, Silver Spring, MD
    1. Centers for Disease Control and Prevention 2008. Syphilis testing algorithms using treponemal tests for initial screening—four laboratories, New York City, 2005-2006. MMWR Morb. Mortal. Wkly. Rep. 57:872–875 - PubMed
    1. Gomez E., Jespersen D. J., Harring J. A., Binnicker M. J. 2010. Evaluation of the Bio-Rad BioPlex 2200 syphilis multiplex flow immunoassay for the detection of IgM- and IgG-class antitreponemal antibodies. Clin. Vaccine Immunol. 17:966–968 - PMC - PubMed
    1. Landis J. R., Koch G. G. 1977. The measurement of observer agreement for categorical data. Biometrics 33:159–174 - PubMed
    1. Larsen S. A., Steiner B. M., Rudolph A. H. 1995. Laboratory diagnosis and interpretation of tests for syphilis. Clin. Microbiol. Rev. 8:1–21 - PMC - PubMed

Substances