Wrong intraocular lens implant; learning from reported patient safety incidents

Affiliations

Wrong intraocular lens implant; learning from reported patient safety incidents

S P Kelly et al. Eye (Lond). 2011 Jun.

Abstract

Purpose: To consider wrong intraocular lens (IOL) implant events in cataract surgical care reported through a national incident reporting database. To propose potential solutions for such events where possible.

Methods: Thematic retrospective review of wrong IOL implantation incidents, as reported through clinical incident reporting methods in NHS care in England and Wales from 2003 to 2010, ascertained from database mining at the National Patient Safety Agency.

Results: In total, 164 patient safety incident (PSI) reports of wrong IOL implantation were located from the study period and considered. There were 47 reports where further surgical intervention was required. All, but one of these required IOL exchange surgery. A total of 62 reports did not provide any causal reason for the wrong IOL implantation and thus provide little if any potential learning. Inaccurate biometry (n=29), wrong IOL selection (n=21), transcription errors (n=10) and handwriting misinterpretations (n=7) were causal reasons reported and are thus potential areas for ophthalmic teams to review and improve practice.

Conclusion: Although infrequent, biometry/IOL implant errors or wrong implants do occasionally occur during cataract care and are thus a threat to quality. There is room for improvement in incident reporting in NHS cataract care as root causation of error was usually lacking in the PSI reports. Nevertheless, lessons for improvement of care from a national incident reporting database for a frequently undertaken surgical procedure were found. Suggestions are proposed for improving quality by reducing wrong IOL problems in cataract care based on analysis of such reports.

PubMed Disclaimer

References

    1. The Information Centre, hospital episode statistics, hesonline Main procedures and interventions: 2000–2009 . http://www.hesonline.nhs.uk/Ea se/servlet/ContentServer?siteID=1937&cate... .
    1. Simon JW, Ngo Y, Khan S, Strogatz D. Surgical confusions in ophthalmology. Arch Ophthalmol. 2007;125:1515–1522. - PubMed
    1. Ali N.A decade of clinical negligence in ophthalmology BMC Ophthalmol 2007720doi: 10.1186/1471-2415-7-20 - DOI - PMC - PubMed
    1. Ali N, Little BC.Causes of cataract surgery malpractice claims in England 1995–2008 Br J Ophthalmol 2010. 182774. (e-pub ahead of print 30 August 2010; doi:10.1136/bjo.2010.182774 - DOI - PubMed
    1. Brick DC. Risk management lessons from a review of 168 cataract surgery claims. Surv Ophthalmol. 1999;43 (4:356–360. - PubMed