Comparative Study
. 2011 Aug;51(8):1695-706.
doi: 10.1111/j.1537-2995.2011.03066.x. Epub 2011 Feb 25.

A comparison of fibrinogen measurement methods with fibrin clot elasticity assessed by thromboelastometry, before and after administration of fibrinogen concentrate in cardiac surgery patients

Affiliations
Comparative Study

A comparison of fibrinogen measurement methods with fibrin clot elasticity assessed by thromboelastometry, before and after administration of fibrinogen concentrate in cardiac surgery patients

Cristina Solomon et al. Transfusion. 2011 Aug.

Abstract

Background: Fibrinogen concentrate administration can be guided by measuring fibrinogen concentration or quality of the fibrin-based clot. This study compared different fibrinogen concentration measurement methods with maximum clot firmness (MCF) of the fibrin clot, assessed by thromboelastometry (FIBTEM), in 33 cardiovascular surgery patients receiving fibrinogen concentrate for hemostatic therapy.

Study design and methods: Blood samples were collected after cardiopulmonary bypass (CPB) and after fibrinogen concentrate administration. FIBTEM MCF was measured using a rotational thromboelastometry device (ROTEM, Tem International). Fibrinogen concentration was measured using photo-optical (CA-7000, Siemens Healthcare Diagnostics), mechanical (KC-10 steel ball, Schnitger and Gross hook, Amelung GmbH), and electromechanical (STA-R, Diagnostica Stago) coagulometers. Assessments included agreement between fibrinogen concentration measurements and correlations between fibrinogen concentration and FIBTEM MCF.

Results: After CPB, correlations were significant (p < 0.001) between FIBTEM MCF and fibrinogen concentration determined by steel ball (r = 0.71), hook (r = 0.73), STA-R (r = 0.81), and CA-7000 (r = 0.82) coagulometers. After fibrinogen concentrate administration, agreement between fibrinogen measurement methods was severely impaired, and correlations with FIBTEM MCF were 0.39 (steel ball), 0.33 (hook), 0.59 (STA-R), and 0.33 (CA-7000).

Conclusion: Agreement between fibrinogen concentration measurement methods decreased considerably after fibrinogen concentrate administration. All methods correlated acceptably with FIBTEM MCF at the end of CPB, but not after hemostatic therapy. Further investigation is needed to explain these findings.