Evaluation of an intervention to maintain endotracheal tube cuff pressure within therapeutic range
- PMID: 21362715
- PMCID: PMC3506174
- DOI: 10.4037/ajcc2011661
Evaluation of an intervention to maintain endotracheal tube cuff pressure within therapeutic range
Abstract
Background: Endotracheal tube cuff pressure must be kept within an optimal range that ensures ventilation and prevents aspiration while maintaining tracheal perfusion.
Objectives: To test the effect of an intervention (adding or removing air) on the proportion of time that cuff pressure was between 20 and 30 cm H(2)O and to evaluate changes in cuff pressure over time.
Methods: A repeated-measure crossover design was used to study 32 orally intubated patients receiving mechanical ventilation for two 12-hour shifts (randomized control and intervention conditions). Continuous cuff pressure monitoring was initiated, and the pressure was adjusted to a minimum of 22 cm H(2)O. Caregivers were blinded to cuff pressure data, and usual care was provided during the control condition. During the intervention condition, cuff pressure alarm or clinical triggers guided the intervention.
Results: Most patients were men (mean age, 61.6 years). During the control condition, 51.7% of cuff pressure values were out of range compared with 11.1% during the intervention condition (P < .001). During the intervention, a mean of 8 adjustments were required, mostly to add air to the endotracheal tube cuff (mean 0.28 [SD, 0.13] mL). During the control condition, cuff pressure decreased over time (P < .001).
Conclusions: The intervention was effective in maintaining cuff pressure within an optimal range, and cuff pressure decreased over time without intervention. The effect of the intervention on outcomes such as ventilator-associated pneumonia and tracheal damage requires further study.
Figures
Comment in
-
Deviations in endotracheal cuff pressure during intensive care.Am J Crit Care. 2011 Nov;20(6):421-2; author reply 422. doi: 10.4037/ajcc2011398. Am J Crit Care. 2011. PMID: 22045130 No abstract available.
-
Endotracheal cuff pressure monitoring: another alarm in the ICU?Am J Crit Care. 2011 Nov;20(6):422-3; author reply 423-4. doi: 10.4037/ajcc2011725. Am J Crit Care. 2011. PMID: 22045132 No abstract available.
References
-
- Diaz E, Rodriguez AH, Rello J. Ventilator-associated pneumonia: issues related to the artificial airway. Respir Care. 2005;50:900–906. - PubMed
-
- Rello J, Sonora R, Jubert P, Artigas A, Rue M, Valles J. Pneumonia in intubated patients: role of respiratory airway care. Am J Respir Crit Care Med. 1996;154:111–115. - PubMed
-
- Simmons KF, Scanlan CL. Airway management. In: Wilkins RL, Stoller JK, Scanlan CL, editors. Egan’s Fundamentals of Respiratory Care. 8. St Louis, MO: Mosby; 2003.
-
- Guidelines for preventing health-care-associated pneumonia, 2003 recommendations of the CDC the Healthcare Infection Control Practices Advisory Committee. Respir Care. 2004;49:926–939. - PubMed
-
- Crimlisk JT, Horn MH, Wilson DJ, Marino B. Artificial airways: a survey of cuff management practices. Heart Lung. 1996;25:225–235. - PubMed