. 2011 Jun;26(6):799-808.
doi: 10.1007/s00384-011-1175-y. Epub 2011 Mar 15.

In vivo validation of a therapy planning system for laser-induced thermotherapy (LITT) of liver malignancies

Affiliations

In vivo validation of a therapy planning system for laser-induced thermotherapy (LITT) of liver malignancies

Kai Siegfried Lehmann et al. Int J Colorectal Dis. 2011 Jun.

Abstract

Purpose: In situ ablation is increasingly being used for the treatment of liver malignancies. The application of these techniques is limited by the lack of a precise prediction of the destruction volume. This holds especially true in anatomically difficult situations, such as metastases in the vicinity of larger liver vessels. We developed a three-dimensional (3D) planning system for laser-induced thermotherapy (LITT) of liver tumors. The aim of the study was to validate the system for calculation of the destruction volume.

Methods: LITT (28 W, 20 min) was performed in close contact to major hepatic vessels in six pigs. After explantation of the liver, the coagulation area was documented. The liver and its vascular structures were segmented from a pre-interventional CT scan. Therapy planning was carried out including the cooling effect of adjacent liver vessels. The lesions in vivo and the simulated lesions were compared with a morphometric analysis.

Results: The volume of lesions in vivo was 6,568.3 ± 3,245.9 mm(3), which was not different to the simulation result of 6,935.2 ± 2,538.5 mm(3) (P = 0.937). The morphometric analysis showed a sensitivity of the system of 0.896 ± 0.093 (correct prediction of destructed tissue). The specificity was 0.858 ± 0.090 (correct prediction of vital tissue).

Conclusions: A 3D computer planning system for the prediction of thermal lesions in LITT was developed. The calculation of the directional cooling effect of intrahepatic vessels is possible for the first time. The morphometric analysis showed a good correlation under clinical conditions. The pre-therapeutic calculation of the ablation zone might be a valuable tool for procedure planning.

PubMed Disclaimer