Effect of +3.00 diopter and +4.00 diopter additions in multifocal intraocular lenses on defocus profiles, patient satisfaction, and contrast sensitivity
- PMID: 21420598
- DOI: 10.1016/j.jcrs.2010.11.027
Effect of +3.00 diopter and +4.00 diopter additions in multifocal intraocular lenses on defocus profiles, patient satisfaction, and contrast sensitivity
Abstract
Purpose: To evaluate the advantages and disadvantages of the new low-addition (add) (+3.00 diopter [D]) ReSTOR multifocal IOL compared with the preceding ReSTOR model with +4.00 D add.
Setting: University Eye Hospital, Tuebingen, Germany.
Design: Comparative case series.
Methods: Patients with a +3.00 D or +4.00 D add multifocal IOL were examined for uncorrected and distance-corrected visual acuity at distance, intermediate, and near. A defocus profile was assessed, individual reading distance and the distance for lowest intermediate visual acuity were determined. Patient satisfaction was evaluated with a standardized questionnaire. Contrast sensitivity was tested under mesopic and photopic conditions.
Results: Uncorrected and distance-corrected intermediate visual acuities were statistically significantly better in the +3.00 D add group (24 eyes) than in the +4.00 D add group (30 eyes); distance and near visual acuities were not different between groups. The defocus profile significantly varied between groups. The +4.00 D add group had a closer reading distance (33.0 cm) than the +3.00 D add group (43.5 cm), a closer point of lowest intermediate visual acuity (65.8 cm versus 86.9 cm) and worse lowest intermediate visual acuity (20/59 ± 4.5 letters [SD] versus 20/48 ± 5.5 letters). Thus, patients in the +3.00 D add group reported being more satisfied with intermediate visual acuity. The +3.00 D add group reported more glare but less halos than the +4.00 D add group; contrast sensitivity was not different.
Conclusion: The lower addition resulted in a narrower defocus profile, a farther reading distance, and better intermediate visual acuity and thus increased patient satisfaction.
Financial disclosure: No author has a financial or proprietary interest in any material or method mentioned. Additional disclosure is found in the footnotes.