Measuring peak expiratory flow in general practice: comparison of mini Wright peak flow meter and turbine spirometer

Affiliations
Clinical Trial

Measuring peak expiratory flow in general practice: comparison of mini Wright peak flow meter and turbine spirometer

K P Jones et al. BMJ. .

Abstract

Objective: To compare measurements of the peak expiratory flow rate taken by the mini Wright peak flow meter and the turbine spirometer.

Design: Pragmatic study with randomised order of use of recording instruments. Phase 1 compared a peak expiratory flow type expiration recorded by the mini Wright peak flow meter with an expiration to forced vital capacity recorded by the turbine spirometer. Phase 2 compared peak expiratory flow type expirations recorded by both meters. Reproducibility was assessed separately.

Setting: Routine surgeries at Aldermoor Health Centre, Southampton.

Subjects: 212 Patients aged 4 to 78 presenting with asthma or obstructive airways disease. Each patient contributed only once to each phase (105 in phase 1, 107 in phase 2), but some entered both phases on separate occasions. Reproducibility was tested on a further 31 patients.

Main outcome measure: 95% Limits of agreement between measurements on the two meters.

Results: 208 (98%) Of the readings taken by the mini Wright meter were higher than the corresponding readings taken by the turbine spirometer, but the 95% limits of agreement (mean difference (2 SD] were wide (1 to 173 l/min). Differences due to errors in reproducibility were not sufficient to predict this level of disagreement. Analysis by age, sex, order of use, and the type of expiration did not detect any significant differences.

Conclusions: The two methods of measuring peak expiratory flow rate were not comparable. The mini Wright meter is likely to remain the preferred instrument in general practice.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Anaesthesia. 1988 Mar;43(3):233-8 - PubMed
    1. Br Med J. 1978 Dec 9;2(6152):1627-8 - PubMed
    1. Thorax. 1979 Feb;34(1):79-81 - PubMed
    1. Thorax. 1987 Sep;42(9):689-93 - PubMed
    1. Lancet. 1986 Feb 8;1(8476):307-10 - PubMed

Publication types

LinkOut - more resources