Revisions of metal-on-metal hip resurfacing: lessons learned and improved outcome
- PMID: 21435500
- DOI: 10.1016/j.ocl.2011.01.003
Revisions of metal-on-metal hip resurfacing: lessons learned and improved outcome
Abstract
This retrospective, consecutive case series of a single surgeon performed between 2001 and 2010 assesses the outcome following revision of metal-on-metal hip resurfacing arthroplasties (N = 113). Mean time to revision was 31 months (0-101) after primary hip resurfacing. Malpositioning of the components with associated wear-induced soft tissue fluid collections was the most frequent factor leading to failure of a hip resurfacing arthroplasty. The mid-term outcome of the revisions was satisfactory; complications occurred in 11 patients (9.7%). Six of these patients underwent a re-revision.
Copyright © 2010 Elsevier Ltd. All rights reserved.