Impact of tube current-time and tube voltage reduction in 64-detector-row computed tomography pulmonary angiography for pulmonary embolism in a porcine model
- PMID: 21436746
- DOI: 10.1097/RTI.0b013e31820b8696
Impact of tube current-time and tube voltage reduction in 64-detector-row computed tomography pulmonary angiography for pulmonary embolism in a porcine model
Abstract
Purpose: To evaluate the impact of dose reduction in multidetector computed tomography pulmonary angiography (CTPA) for detection of pulmonary embolism (PE).
Materials and methods: After induction of PE in 6 anesthetized pigs, a 64-detector-row CTPA was performed at 3 different dose protocols: A (120 kV/120 mAseff.), B (120 kV/80 mAseff.), and C (80 kV/80 mAseff.). Images were evaluated by 2 radiologists independently. A high-dose CTPA (120 kV/250 mAseff.) served as a reference standard. Sensitivity, specificity, and positive and negative predictive values were calculated and compared using the Wilcoxon test. Interobserver agreement was determined by calculation of κ values. Radiation exposure and objective image parameters were assessed and compared with a 2-sided t test.
Results: In the reference scan, a total of 94 emboli were detected: 17 in the main and lobar pulmonary level (category A), 47 in the segmental level (category B), and 30 in the subsegmental level (category C). All protocols reached high diagnostic accuracy in the detection of PE in category A. No significant difference was observed between protocols A and B in the detection rate of segmental and subsegmental PE (sensitivity: 93.6% and 91.5% vs 85.1% and 87.2%; positive predictive value: 100% and 97.7% vs 97.5% and 95.3%). Interobserver agreement was excellent at the segmental (κ=0.97 and 0.94) and subsegmental levels (κ=0.94 and 0.92). Using protocol C, the detection rate of segmental and subsegmental emboli was significantly impaired and interobserver agreement was significantly inferior (sensitivity: 46.8% and 44.6% at segmental and 56.7% and 50.0% at subsegmental level; κ=0.70 and 0.60, respectively; P<0.05). Vessel attenuation was significantly higher in protocol C, [710.3 Hounsfield units (HU) ±200.6 SD] than in protocols A (414.4 HU±82.5 SD) and B (428.8 HU±78.9 SD) (P<0.001). Signal-to-noise and contrast-to-noise ratios were significantly decreased in protocols B (55.6; 45.8) and C (44.3; 39.4) compared with protocol A (62.1; 52.3) (P<0.05). Radiation dose was significantly reduced between protocols A and B and protocols B and C [volume CT dose index (CTDIvol): 7.1 vs 4.7 vs 1.4 mGy; dose-length product: 199.5 vs 132.1 vs 39.4 mGy·cm and E: 3.39 vs 2.25 vs 0.6 mSv; P<0.05].
Conclusion: This animal study proves the diagnostic use of a low-dose CTPA protocol with 80 mAseff. at 120 kV for PE evaluation without significant loss of diagnostic accuracy compared with a standard protocol with 120 mAseff. Using an "ultra-low"-dose CTPA protocol (80 kV; 80 mAseff.), emboli detection at the segmental and subsegmental level is significantly impaired, whereas evaluation of central emboli is still feasible with high diagnostic accuracy.
Similar articles
-
Iterative reconstructed ultra high pitch CT pulmonary angiography with cardiac bowtie-shaped filter in the acute setting: effect on dose and image quality.Eur J Radiol. 2013 Sep;82(9):1571-6. doi: 10.1016/j.ejrad.2013.04.043. Epub 2013 Jun 15. Eur J Radiol. 2013. PMID: 23777744 Clinical Trial.
-
Intraindividual comparison of gadolinium- and iodine-enhanced 64-slice multidetector CT pulmonary angiography for the detection of pulmonary embolism in a porcine model.Emerg Radiol. 2011 Jun;18(3):189-95. doi: 10.1007/s10140-010-0926-3. Epub 2010 Dec 7. Emerg Radiol. 2011. PMID: 21136130
-
Improved image quality and low radiation dose with hybrid iterative reconstruction with 80 kV CT pulmonary angiography.Eur J Radiol. 2014 Oct;83(10):1962-9. doi: 10.1016/j.ejrad.2014.06.016. Epub 2014 Jul 8. Eur J Radiol. 2014. PMID: 25084687
-
Peripheral pulmonary embolism on multidetector CT pulmonary angiography.JBR-BTR. 2007 Mar-Apr;90(2):100-8. JBR-BTR. 2007. PMID: 17555069 Review.
-
Iterative reconstruction of pulmonary MDCT angiography: effects on image quality, effective dose and estimated organ dose to the breast.Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2014 Jun;158(2):259-64. doi: 10.5507/bp.2013.059. Epub 2013 Sep 12. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2014. PMID: 24026145 Review.
Cited by
-
Evaluation of image quality and radiation dose saving comparing knowledge model-based iterative reconstruction on 80-kV CT pulmonary angiography (CTPA) with hybrid iterative reconstruction on 100-kV CT.Emerg Radiol. 2019 Apr;26(2):145-153. doi: 10.1007/s10140-018-1653-4. Epub 2018 Nov 10. Emerg Radiol. 2019. PMID: 30415416
-
Image quality of low mA CT pulmonary angiography reconstructed with model based iterative reconstruction versus standard CT pulmonary angiography reconstructed with filtered back projection: an equivalency trial.Eur Radiol. 2015 Jun;25(6):1665-71. doi: 10.1007/s00330-014-3563-5. Epub 2014 Dec 21. Eur Radiol. 2015. PMID: 25528671
-
Emphysema quantification using hybrid versus model-based generations of iterative reconstruction: SAFIRE versus ADMIRE.Medicine (Baltimore). 2019 Feb;98(7):e14450. doi: 10.1097/MD.0000000000014450. Medicine (Baltimore). 2019. PMID: 30762757 Free PMC article.
-
Ultra Low Dose CT Pulmonary Angiography with Iterative Reconstruction.PLoS One. 2016 Sep 9;11(9):e0162716. doi: 10.1371/journal.pone.0162716. eCollection 2016. PLoS One. 2016. PMID: 27611830 Free PMC article.
-
CT Pulmonary Angiography at Reduced Radiation Exposure and Contrast Material Volume Using Iterative Model Reconstruction and iDose4 Technique in Comparison to FBP.PLoS One. 2016 Sep 9;11(9):e0162429. doi: 10.1371/journal.pone.0162429. eCollection 2016. PLoS One. 2016. PMID: 27611448 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical