Comparison of high sensitivity and contemporary troponin assays for the early detection of acute myocardial infarction in the emergency department
- PMID: 21441390
- DOI: 10.1258/acb.2010.010219
Comparison of high sensitivity and contemporary troponin assays for the early detection of acute myocardial infarction in the emergency department
Erratum in
- Ann Clin Biochem. 2012 Mar;49(Pt 2):208. Flaws, Dylan Finlay [added]; Borowsky, Jennifer [added]
Abstract
Background: Current guidelines define acute myocardial infarction (AMI) by the rise and/or fall of cardiac troponin with ≥1 value above the 99th percentile. Past troponin assays have been unreliable at the lower end of the range. Highly sensitive assays have therefore been developed to increase the clinical sensitivity for detection of myocardial injury.
Methods: Three hundred and thirty-two patients with chest pain suggestive of AMI were prospectively recruited between November 2006 and April 2007. Serial blood samples were analysed to compare Roche Elecsys high sensitivity troponin T (hsTnT), Abbott Architect troponin I 3rd generation (TnI 3) and Roche Elecsys troponin T (TnT) for the diagnosis of AMI.
Results: One hundred and ten (33.1%) patients were diagnosed with AMI. Test performance for the diagnosis of AMI, as quantified by receiver operating characteristic area under the curve (95% confidence intervals) for baseline/follow-up troponins were as follows: hsTnT 0.90 (0.87-0.94)/0.94 (0.91-0.97), TnI 3 0.88 (0.84-0.92)/0.93 (0.90-0.96) and TnT 0.80 (0.74-0.85)/0.89 (0.85-0.94). hsTnT was superior to TnT (P < 0.001/0.013 at baseline/follow-up) but equivalent to TnI 3. For patients with a final diagnosis of AMI, baseline troponins were raised in more patients for hsTnT (83.6%) than TnI 3 (74.5%) and TnT (62.7%). A delta troponin of ≥20% increased the specificity of hsTnT from 80.6% to 93.7% but reduced sensitivity from 90.9% to 71.8%.
Conclusion: hsTnT was superior to TnT but equivalent to TnI 3 for the diagnosis of AMI. Serial troponin measurement increased test performance. hsTnT was the most likely to be raised at baseline in those with AMI. A delta troponin increases specificity but reduces sensitivity.
Comment in
-
Comparison of high sensitivity and contemporary troponin I immunoassays for the early detection of acute myocardial infarction in the emergency department.Ann Clin Biochem. 2012 Mar;49(Pt 2):205-6. doi: 10.1258/acb.2011.011211. Epub 2012 Feb 9. Ann Clin Biochem. 2012. PMID: 22323661 No abstract available.
-
High-sensitivity troponin assays and clinical decisions.Ann Clin Biochem. 2012 May;49(Pt 3):306-7; author reply 307. doi: 10.1258/acb.2011.011137. Epub 2012 Mar 15. Ann Clin Biochem. 2012. PMID: 22422152 No abstract available.
Similar articles
-
High sensitivity troponin outperforms contemporary assays in predicting major adverse cardiac events up to two years in patients with chest pain.Ann Clin Biochem. 2011 May;48(Pt 3):249-55. doi: 10.1258/acb.2010.010220. Epub 2011 Mar 25. Ann Clin Biochem. 2011. PMID: 21441393
-
Early diagnosis of acute myocardial infarction in the elderly using more sensitive cardiac troponin assays.Eur Heart J. 2011 Jun;32(11):1379-89. doi: 10.1093/eurheartj/ehr033. Epub 2011 Feb 28. Eur Heart J. 2011. PMID: 21362702
-
High-sensitivity troponin T for early rule-out of myocardial infarction in recent onset chest pain.Emerg Med J. 2012 Oct;29(10):805-10. doi: 10.1136/emermed-2011-200222. Epub 2011 Nov 22. Emerg Med J. 2012. PMID: 22109535
-
[Biochemical marker for myocardial injury].Rinsho Byori. 2002 Dec;50(12):1140-5. Rinsho Byori. 2002. PMID: 12652682 Review. Japanese.
-
Troponins and other biomarkers in the early diagnosis of acute myocardial infarction.Postgrad Med J. 2015 Jun;91(1076):322-30. doi: 10.1136/postgradmedj-2014-133129. Epub 2015 May 22. Postgrad Med J. 2015. PMID: 26002926 Review.
Cited by
-
High-Sensitivity Troponin I and Cardiovascular Events in Stable Coronary Artery Disease: Insights from a Longitudinal Outpatient Study.Int J Mol Sci. 2023 Dec 9;24(24):17286. doi: 10.3390/ijms242417286. Int J Mol Sci. 2023. PMID: 38139114 Free PMC article.
-
Diagnostic accuracy of sensitive or high-sensitive troponin on presentation for myocardial infarction: a meta-analysis and systematic review.Vasc Health Risk Manag. 2014 Jul 21;10:435-50. doi: 10.2147/VHRM.S63416. eCollection 2014. Vasc Health Risk Manag. 2014. PMID: 25092986 Free PMC article.
-
Trimethyllysine, a trimethylamine N-oxide precursor, provides near- and long-term prognostic value in patients presenting with acute coronary syndromes.Eur Heart J. 2019 Aug 21;40(32):2700-2709. doi: 10.1093/eurheartj/ehz259. Eur Heart J. 2019. PMID: 31049589 Free PMC article.
-
Serial high-sensitivity cardiac troponin testing for the diagnosis of myocardial infarction: a scoping review.BMJ Open. 2022 Nov 22;12(11):e066429. doi: 10.1136/bmjopen-2022-066429. BMJ Open. 2022. PMID: 36414302 Free PMC article.
-
Diagnostic accuracy of single baseline measurement of Elecsys Troponin T high-sensitive assay for diagnosis of acute myocardial infarction in emergency department: systematic review and meta-analysis.BMJ. 2015 Jan 20;350:h15. doi: 10.1136/bmj.h15. BMJ. 2015. PMID: 25646632 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous