. 2011 May;106(5):617-31.
doi: 10.1093/bja/aer058. Epub 2011 Mar 29.

Major complications of airway management in the UK: results of the Fourth National Audit Project of the Royal College of Anaesthetists and the Difficult Airway Society. Part 1: anaesthesia

Affiliations
Free article

Major complications of airway management in the UK: results of the Fourth National Audit Project of the Royal College of Anaesthetists and the Difficult Airway Society. Part 1: anaesthesia

T M Cook et al. Br J Anaesth. 2011 May.
Free article

Abstract

Background: This project was devised to estimate the incidence of major complications of airway management during anaesthesia in the UK and to study these events.

Methods: Reports of major airway management complications during anaesthesia (death, brain damage, emergency surgical airway, unanticipated intensive care unit admission) were collected from all National Health Service hospitals for 1 yr. An expert panel assessed inclusion criteria, outcome, and airway management. A matched concurrent census estimated a denominator of 2.9 million general anaesthetics annually.

Results: Of 184 reports meeting inclusion criteria, 133 related to general anaesthesia: 46 events per million general anaesthetics [95% confidence interval (CI) 38-54] or one per 22,000 (95% CI 1 per 26-18,000). Anaesthesia events led to 16 deaths and three episodes of persistent brain damage: a mortality rate of 5.6 per million general anaesthetics (95% CI 2.8-8.3): one per 180,000 (95% CI 1 per 352-120,000). These estimates assume that all such cases were captured. Rates of death and brain damage for different airway devices (facemask, supraglottic airway, tracheal tube) varied little. Airway management was considered good in 19% of assessable anaesthesia cases. Elements of care were judged poor in three-quarters: in only three deaths was airway management considered exclusively good.

Conclusions: Although these data suggest the incidence of death and brain damage from airway management during general anaesthesia is low, statistical analysis of the distribution of reports suggests as few as 25% of relevant incidents may have been reported. It therefore provides an indication of the lower limit for incidence of such complications. The review of airway management indicates that in a majority of cases, there is 'room for improvement'.

PubMed Disclaimer

Comment in

Publication types

MeSH terms