Perforation of the IVC: rule rather than exception after longer indwelling times for the Günther Tulip and Celect retrievable filters
- PMID: 21448771
- DOI: 10.1007/s00270-011-0151-9
Perforation of the IVC: rule rather than exception after longer indwelling times for the Günther Tulip and Celect retrievable filters
Abstract
Purpose: This study was designed to assess the incidence, magnitude, and impact upon retrievability of vena caval perforation by Günther Tulip and Celect conical inferior vena cava (IVC) filters on computed tomographic (CT) imaging.
Methods: Günther Tulip and Celect IVC filters placed between July 2007 and May 2009 were identified from medical records. Of 272 IVC filters placed, 50 (23 Günther Tulip, 46%; 27 Celect, 54%) were retrospectively assessed on follow-up abdominal CT scans performed for reasons unrelated to the filter. Computed tomography scans were examined for evidence of filter perforation through the vena caval wall, tilt, or pericaval tissue injury. Procedure records were reviewed to determine whether IVC filter retrieval was attempted and successful.
Results: Perforation of at least one filter component through the IVC was observed in 43 of 50 (86%) filters on CT scans obtained between 1 and 880 days after filter placement. All filters imaged after 71 days showed some degree of vena caval perforation, often as a progressive process. Filter tilt was seen in 20 of 50 (40%) filters, and all tilted filters also demonstrated vena caval perforation. Transjugular removal was attempted in 12 of 50 (24%) filters and was successful in 11 of 12 (92%).
Conclusions: Longer indwelling times usually result in vena caval perforation by retrievable Günther Tulip and Celect IVC filters. Although infrequently reported in the literature, clinical sequelae from IVC filter components breaching the vena cava can be significant. We advocate filter retrieval as early as clinically indicated and increased attention to the appearance of IVC filters on all follow-up imaging studies.
Similar articles
-
Caval penetration by retrievable inferior vena cava filters: a retrospective comparison of Option and Günther Tulip filters.J Vasc Interv Radiol. 2013 Apr;24(4):566-71. doi: 10.1016/j.jvir.2012.12.024. Epub 2013 Feb 23. J Vasc Interv Radiol. 2013. PMID: 23462065
-
Complications of Celect, Günther tulip, and Greenfield inferior vena cava filters on CT follow-up: a single-institution experience.J Vasc Interv Radiol. 2013 Nov;24(11):1723-9. doi: 10.1016/j.jvir.2013.07.023. Epub 2013 Sep 13. J Vasc Interv Radiol. 2013. PMID: 24041915
-
Günther Tulip and Celect IVC filters in multiple-trauma patients.J Endovasc Ther. 2009 Aug;16(4):494-9. doi: 10.1583/09-2728.1. J Endovasc Ther. 2009. PMID: 19702346
-
Inferior Vena Cava Filter Long Term Complications and Retrieval Techniques: A Case Series and Literature Review.Vasc Endovascular Surg. 2024 Jul;58(5):559-566. doi: 10.1177/15385744231226048. Epub 2024 Jan 9. Vasc Endovascular Surg. 2024. PMID: 38196287 Review.
-
Filter-Associated Inferior Vena Cava Thrombosis with Duodenal Perforation: Case Report and Literature Review.Ann Vasc Surg. 2019 Jul;58:383.e1-383.e6. doi: 10.1016/j.avsg.2018.11.021. Epub 2019 Feb 11. Ann Vasc Surg. 2019. PMID: 30763706 Review.
Cited by
-
Retrievable inferior vena cava filters can be placed and removed with a high degree of success: Initial experience.Catheter Cardiovasc Interv. 2015 Oct;86(4):719-25. doi: 10.1002/ccd.25716. Epub 2015 Aug 10. Catheter Cardiovasc Interv. 2015. PMID: 25367646 Free PMC article.
-
Vena Cava Filters: Toward Optimal Strategies for Filter Retrieval and Patients' Follow-Up.Front Cardiovasc Med. 2022 Mar 3;9:746748. doi: 10.3389/fcvm.2022.746748. eCollection 2022. Front Cardiovasc Med. 2022. PMID: 35310979 Free PMC article. Review.
-
A Dedicated Inferior Vena Cava Filter Service Line: How to Optimize Your Practice.Semin Intervent Radiol. 2016 Jun;33(2):105-8. doi: 10.1055/s-0036-1582122. Semin Intervent Radiol. 2016. PMID: 27247479 Free PMC article. Review.
-
Creation of an iOS and Android Mobile Application for Inferior Vena Cava (IVC) Filters: A Powerful Tool to Optimize Care of Patients with IVC Filters.Semin Intervent Radiol. 2016 Jun;33(2):137-43. doi: 10.1055/s-0036-1583206. Semin Intervent Radiol. 2016. PMID: 27247483 Free PMC article. Review.
-
Inferior vena cava filter use and patient safety: legacy or science?Hematology Am Soc Hematol Educ Program. 2017 Dec 8;2017(1):686-692. doi: 10.1182/asheducation-2017.1.686. Hematology Am Soc Hematol Educ Program. 2017. PMID: 29222322 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources