A pictographic atlas for classifying damage modes on polyethylene bearings
- PMID: 21461698
- DOI: 10.1007/s10856-011-4303-x
A pictographic atlas for classifying damage modes on polyethylene bearings
Abstract
Evaluation of medical devices retrieved after in vivo service provides unique evidence related to the physiological environment in which the biomaterials performed. This study implements a training procedure for evaluating polyethylene bearings of joint prostheses obtained after pre-clinical tests or explanted after in vivo function. A total of 161 damage regions on 45 bearings were evaluated by four observers. An illustrated Damage Mode Atlas was developed as a reference guide, inclusive of both photographs and concise written descriptions of 16 specific damage modes that are typical for polyethylene bearings. Utilizing the Damage Mode Atlas to train new researchers improved the damage pattern analysis, including more accurate identification of damage modes and improved inter-rater reliability. This Damage Mode Atlas is a useful supplementary tool for conducting Stage II non-destructive analysis of explanted polyethylene bearings used for joint replacement, in accordance with international guidelines for evaluating explanted medical devices.
Similar articles
-
Analysis of wear asymmetry in a series of 94 retrieved polyethylene tibial bearings.J Biomech. 2005 Feb;38(2):367-75. doi: 10.1016/j.jbiomech.2004.02.016. J Biomech. 2005. PMID: 15598465
-
Polyethylene insert damage in unicondylar knee replacement: a comparison of in vivo function and in vitro simulation.Proc Inst Mech Eng H. 2010;224(7):823-30. doi: 10.1243/09544119JEIM699. Proc Inst Mech Eng H. 2010. PMID: 20839650
-
Factors associated with the loss of thickness of polyethylene tibial bearings after knee arthroplasty.J Bone Joint Surg Am. 2007 Jun;89(6):1306-14. doi: 10.2106/JBJS.F.00667. J Bone Joint Surg Am. 2007. PMID: 17545435
-
Ceramic bearings in total knee arthroplasty.J Knee Surg. 2007 Oct;20(4):261-70. doi: 10.1055/s-0030-1248055. J Knee Surg. 2007. PMID: 17993065 Review.
-
[Implant with a mobile or a fixed bearing in unicompartmental knee joint replacemen].Orthopade. 2007 Dec;36(12):1106-12. doi: 10.1007/s00132-007-1159-3. Orthopade. 2007. PMID: 17955209 Review. German.
Cited by
-
A rare case of bilateral large osteolysis following cemented and cementless total knee arthroplasties.Acta Orthop. 2013 Feb;84(1):112-5. doi: 10.3109/17453674.2013.752693. Epub 2012 Dec 17. Acta Orthop. 2013. PMID: 23244786 Free PMC article. No abstract available.
-
Chemical, Mechanical and Tribological Effects of Artificially Aging up to 6 Weeks on Virgin and Crosslinked UHMWPE Evaluated for a TKR Design.Bioengineering (Basel). 2025 Jul 24;12(8):793. doi: 10.3390/bioengineering12080793. Bioengineering (Basel). 2025. PMID: 40868307 Free PMC article.
-
Peripheral snap-fit locking mechanisms and smooth surface finish of tibial trays reduce backside wear in fixed-bearing total knee arthroplasty.Acta Orthop. 2017 Feb;88(1):62-69. doi: 10.1080/17453674.2016.1248202. Epub 2016 Oct 26. Acta Orthop. 2017. PMID: 27781667 Free PMC article.
-
Validation of a model-based measurement of the minimum insert thickness of knee prostheses: a retrieval study.Bone Joint Res. 2014 Oct;3(10):289-96. doi: 10.1302/2046-3758.310.2000304. Bone Joint Res. 2014. PMID: 25278502 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources