Migration pattern of cementless press fit cups in the presence of stabilizing screws in total hip arthroplasty
- PMID: 21486725
- PMCID: PMC3352209
- DOI: 10.1186/2047-783x-16-3-127
Migration pattern of cementless press fit cups in the presence of stabilizing screws in total hip arthroplasty
Abstract
The aim of this study was to evaluate the initial acetabular implant stability and late acetabular implant migration in press fit cups combined with screw fixation of the acetabular component in order to answer the question whether screws are necessary for the fixation of the acetabular component in cementless primary total hip arthroplasty. One hundred and seven hips were available for follow-up after primary THA using a cementless, porous-coated acetabular component. A total of 631 standardized radiographs were analyzed digitally by the "single-film-x-ray-analysis" method (EBRA). One hundred and one (94.4 %) acetabular components did not show significant migration of more than 1 mm. Six (5.6%) implants showed migration of more than 1 mm. Statistical analysis did not reveal preoperative patterns that would identify predictors for future migration. Our findings suggest that the use of screw fixation for cementless porous-coated acetabular components for primary THA does not prevent cup migration.
Figures
References
-
- Dorr LD, Wan Z, Cohen J. Hemispheric titanium porous coated acetabular component without screw fixation. Clin Orthop Relat Res. 1998;35(1):158–68. - PubMed
-
- Valle AG, Zoppi A, Peterson MG, Salvati EA. Clinical and radiographic results associated with a modern, cementless modular cup design in total hip arthroplasty. J Bone Joint Surg Am. 2004;86-A(9):1998–2004. - PubMed
-
- Engh CA, Hopper RH Jr, Engh CA Jr. Long-term porous-coated cup survivorship using spikes, screws, and press-fitting for initial fixation. J Arthroplasty. 2004;19(7 SupplV):54–60. - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Medical