Comparative study of two laser Doppler blood flowmeters
- PMID: 2149392
- DOI: 10.3109/03091909009015431
Comparative study of two laser Doppler blood flowmeters
Abstract
A new infrared laser Doppler blood flow instrument (Moor MBF3D) was evaluated using an in vitro model allowing measurements over a range of flow velocities and concentrations. The responses correlated well (r = 0.96, p less than 0.01) with those obtained simultaneously using a Perimed PF3 laser Doppler instrument. The different processing bandwidths of the instruments were investigated and the wideband mode of operation is recommended for flow measurements where there may be fast moving red blood cells (rbcs). The infrared instrument is capable of dual-channel operation, and the two channels are shown to respond almost identically for similar changes in blood flow through the in vitro model (r = 0.999, p less than 0.01). The main advantage of the dual-channel instrument is that continuous measurements may be made simultaneously at two different skin sites allowing dynamic flow responses to be compared.
Similar articles
-
Experiments on extracted teeth into the validity of using laser Doppler techniques for recording pulpal blood flow.Arch Oral Biol. 1993 May;38(5):431-9. doi: 10.1016/0003-9969(93)90215-8. Arch Oral Biol. 1993. PMID: 8328924
-
Some aspects of the use of laser Doppler flow meters for recording tissue blood flow.Exp Physiol. 1993 Jan;78(1):1-14. doi: 10.1113/expphysiol.1993.sp003664. Exp Physiol. 1993. PMID: 8448007 Review.
-
In vivo comparison of a twin wavelength laser Doppler flowmeter using He-Ne and laser diode sources.J Med Eng Technol. 1990 May-Jun;14(3):102-10. doi: 10.3109/03091909009015421. J Med Eng Technol. 1990. PMID: 2143783
-
Detection of Site-Specific Blood Flow Variation in Humans during Running by a Wearable Laser Doppler Flowmeter.Sensors (Basel). 2015 Oct 5;15(10):25507-19. doi: 10.3390/s151025507. Sensors (Basel). 2015. PMID: 26445047 Free PMC article. Clinical Trial.
-
Laser-Doppler flowmetry.Crit Rev Biomed Eng. 1990;18(2):125-63. Crit Rev Biomed Eng. 1990. PMID: 2242676 Review.
Cited by
-
Laser Doppler measurements of inner ear blood flow during experimental thrombosis of cochlear blood vessels in the guinea pig.Eur Arch Otorhinolaryngol. 1997;254(2):86-90. doi: 10.1007/BF01526186. Eur Arch Otorhinolaryngol. 1997. PMID: 9065662
-
Review of methodological developments in laser Doppler flowmetry.Lasers Med Sci. 2009 Mar;24(2):269-83. doi: 10.1007/s10103-007-0524-0. Epub 2008 Jan 31. Lasers Med Sci. 2009. PMID: 18236103 Review.
-
Evaluation of laser Doppler imaging to measure blood flow in knee ligaments of adult rabbits.Med Biol Eng Comput. 1996 May;34(3):227-31. doi: 10.1007/BF02520078. Med Biol Eng Comput. 1996. PMID: 8762830
-
Dental pulp testing: a review.Int J Dent. 2009;2009:365785. doi: 10.1155/2009/365785. Epub 2009 Nov 12. Int J Dent. 2009. PMID: 20339575 Free PMC article.
-
Local inhibition of nitric oxide generation in man reduces blood flow in finger pulp but not in hand dorsum skin.J Physiol. 1996 Jan 15;490 ( Pt 2)(Pt 2):501-8. doi: 10.1113/jphysiol.1996.sp021161. J Physiol. 1996. PMID: 8821146 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources