Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers
- PMID: 21504373
- DOI: 10.1515/CCLM.2011.185
Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers
Abstract
Background: The random use of point-of-care, blood gas and core laboratory analyzers to measure electrolytes and metabolites increases the variability in test results. This study was designed to determine whether these results performed on whole blood (point-of-care and blood gas) and plasma (core laboratory) platforms are interchangeable without a risk of clinically relevant discrepancies.
Methods: The interchangeability of the blood gas analysis, electrolytes, glucose, lactate and hemoglobin results performed with three stat platforms (i-STAT, Radiometer ABL 825, RapidLab 865) and two core laboratory platforms (Roche Modular P800 and Sysmex XE-2100) were evaluated using samples from critically ill patients.
Results: For pH, pCO(2), pO(2) and Ca(2+), good correlation (r-values 0.96-1.00) was observed for all comparative analyzers and the biases were within clinically acceptable limits. Potassium, sodium, glucose, lactate and hemoglobin measured with stat analyzers was highly correlated with measurements performed using the laboratory analyzers, r-values 0.89-1.00 and slopes 0.83-1.07. Mean differences with significant bias (p<0.0001) were found for sodium with blood gas analyzers and hemoglobin with i-STAT.
Conclusions: The blood gas, K(+), Na(+), Ca(2+), glucose and lactate results measured with stat and core laboratory analyzers can be used in different clinical settings for critical care management. However, when monitoring small changes in sodium concentrations, the use of single analyzer is encouraged to avoid analytical differences (acceptance limit ± 2%) and misinterpretation of results measured with multiple analyzers. The users of the i-STAT at low hemoglobin values overdiagnose anaemia. Thus, prior to transfusion, the use of hemoglobin concentrations measured with laboratory analyzers is preferable.
Comment in
-
Method comparison: where do we draw the line?Clin Chem Lab Med. 2011 Jul;49(7):1089-90. doi: 10.1515/CCLM.2011.632. Epub 2011 Jun 13. Clin Chem Lab Med. 2011. PMID: 21663561 No abstract available.
Similar articles
-
Comparison of blood gas, electrolyte and metabolite results measured with two different blood gas analyzers and a core laboratory analyzer.Scand J Clin Lab Invest. 2015 Apr;75(2):97-105. doi: 10.3109/00365513.2014.981854. Epub 2014 Nov 28. Scand J Clin Lab Invest. 2015. PMID: 25431133
-
Evaluation of a hand-held blood gas analyzer for rapid determination of blood gases, electrolytes and metabolites in intensive care setting.Clin Chem Lab Med. 2016 Apr;54(4):585-94. doi: 10.1515/cclm-2015-0592. Clin Chem Lab Med. 2016. PMID: 26457784
-
Analysis of bias in measurements of potassium, sodium and hemoglobin by an emergency department-based blood gas analyzer relative to hospital laboratory autoanalyzer results.PLoS One. 2015 Apr 7;10(4):e0122383. doi: 10.1371/journal.pone.0122383. eCollection 2015. PLoS One. 2015. PMID: 25849375 Free PMC article.
-
Compatibility levels between blood gas analysis and central laboratory hemoglobin and electrolyte tests in pediatric patients: A single-center experience.Paediatr Anaesth. 2023 Feb;33(2):107-113. doi: 10.1111/pan.14567. Epub 2022 Oct 10. Paediatr Anaesth. 2023. PMID: 36178754 Review.
-
Point of care hematocrit and hemoglobin in cardiac surgery: a review.Perfusion. 2007 May;22(3):179-83. doi: 10.1177/0267659107080826. Perfusion. 2007. PMID: 18018397 Review.
Cited by
-
Preanalytical conditions of point-of-care testing in the intensive care unit are decisive for analysis reliability.Ann Intensive Care. 2016 Dec;6(1):57. doi: 10.1186/s13613-016-0152-6. Epub 2016 Jun 24. Ann Intensive Care. 2016. PMID: 27342259 Free PMC article.
-
Blood-gas vs. Central-Laboratory analyzers: interchangeability and reference intervals for sodium, potassium, glucose, lactate and hemoglobin.Heliyon. 2021 Oct 30;7(11):e08302. doi: 10.1016/j.heliyon.2021.e08302. eCollection 2021 Nov. Heliyon. 2021. PMID: 34816029 Free PMC article.
-
Bile composition of healthy cats and cats with suspected hepatobiliary disease using point-of-care analyzers: A prospective preliminary study.Can J Vet Res. 2020 Apr;84(2):138-145. Can J Vet Res. 2020. PMID: 32255909 Free PMC article.
-
Hyponatremia with an Osmolar Gap, Pseudohyponatremia or Hyper-Osmolar Hyponatremia?Am J Med Open. 2024 Dec 7;13:100081. doi: 10.1016/j.ajmo.2024.100081. eCollection 2025 Jun. Am J Med Open. 2024. PMID: 39834625 Free PMC article.
-
Plasma/Serum Electrolyte and Metabolite Testing on Blood Gas Analyzer ABL837, a New Application.Diagnostics (Basel). 2025 Jul 31;15(15):1923. doi: 10.3390/diagnostics15151923. Diagnostics (Basel). 2025. PMID: 40804887 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous