Veress needle insertion into the left hypochondrium for creation of pneumoperitoneum: diagnostic value of tests to determine the position of the needle in unselected patients
- PMID: 21537740
- DOI: 10.1590/s0100-69912011000100006
Veress needle insertion into the left hypochondrium for creation of pneumoperitoneum: diagnostic value of tests to determine the position of the needle in unselected patients
Abstract
Objective: To assess the effectiveness of the Veress needle puncture in the left hypochondrium and the accuracy of the tests described for the intraperitoneal correct positioning of the tip of the Veress needle in an unselected population.
Methods: Ninetyone patients consecutively scheduled for Videolaparoscopy had the abdominal wall punctured in the left hypochondrium. There were no exclusion criteria. The patients received general anesthesia and mechanical ventilation according to the protocol. After puncturing five tests were used to confirm the positioning of the needle tip within the peritoneal cavity: aspiration test--AT; resistance to infusion--Pres; recovery of the infused fluid--Prec, dripping test--DT, and test of initial intraperitoneal pressure--IIPP. The test results were compared with results from literature for groups with defined exclusion criteria. The results were used for calculating sensitivity (S) specificity (E), positive predictive value (PPV) and negative predictive value (NPV). Inferential statistical methods were used to analyze the findings.
Results: There were 13 failures. AT had E = 100% and NPV 100%. Pres had S = 100%, E = 0; PPV = 85.71%; NPV does not apply. Prec: S = 100%, E = 53.84%, PPV = 92.85%, NPV = 100%. DT: S = 100%, E = 61.53%, PPV = 93.97% NPV 100%. In IIPP, S, E, PPV and NPV were 100%.
Conclusion: The puncture in the left hypochondrium is effective and the performed tests guide the surgeon regardless of sex, BMI, or previous laparotomy.