Image quality and dose performance of 80 kV low dose scan protocol in high-pitch spiral coronary CT angiography: feasibility study
- PMID: 21541774
- DOI: 10.1007/s10554-011-9822-5
Image quality and dose performance of 80 kV low dose scan protocol in high-pitch spiral coronary CT angiography: feasibility study
Abstract
To investigate the image quality and dose performance of 80 kV high-pitch spiral (HPS) coronary CT angiography (CCTA). 106 patients consecutively enrolled into prospectively ECG-triggering HPS CCTA (pitch = 3.4) exam using kV/ref. mAs = 80/400, 100/370, and 120/370 when patient BMI was ≤22.5 (n = 40), between 22.5 and 27.5 (n = 53) and >27.5 kg/m² (n = 13). Image quality was assessed per-segment by two observers independently using a 4-point scale (1-excellent, 4-non-diagnosable). Image noise and signal-to-noise ratio (SNR), contrast-to-noise ratio were measured. Diagnostic image quality was obtained in 503 of 507, 687 of 693, 164 of 167 coronary segments in 80, 100, 120 kV groups without significant difference (P = 0.482). The proportions of segments with score 1-4 were not significantly different among three kV groups (all P > 0.05). Image noise were significantly higher in 80 kV group than 100 and 120 groups (P < 0.001), while SNR was not (P = 0.097). The effective dose of 80 kV group (0.36 ± 0.03 mSv) was significantly lower than that of 100 kV group (0.86 ± 0.08 mSv) and 120 kV group (1.77 ± 0.18 mSv). The mean ± SD of HR in all patients was 54.8 ± 5.1 bpm. 80 kV HPS CCTA is feasible for patient with BMI ≤ 22.5 kg/m² which can save 58% dose than 100 kV group, while maintain diagnosable image quality.
Comment in
-
Reduction of radiation dose using 80 kV tube voltage: a feasible strategy?Int J Cardiovasc Imaging. 2012 Feb;28(2):425-8. doi: 10.1007/s10554-011-9845-y. Epub 2011 Mar 19. Int J Cardiovasc Imaging. 2012. PMID: 21424152 Free PMC article. No abstract available.
Similar articles
-
320-row CT coronary angiography: effect of 100-kV tube voltages on image quality, contrast volume, and radiation dose.Int J Cardiovasc Imaging. 2011 Oct;27(7):1059-68. doi: 10.1007/s10554-010-9754-5. Epub 2010 Nov 26. Int J Cardiovasc Imaging. 2011. PMID: 21110100
-
Image quality and radiation dose comparison of prospectively triggered low-dose CCTA: 128-slice dual-source high-pitch spiral versus 64-slice single-source sequential acquisition.Int J Cardiovasc Imaging. 2012 Jun;28(5):1217-25. doi: 10.1007/s10554-011-9921-3. Epub 2011 Jul 10. Int J Cardiovasc Imaging. 2012. PMID: 21744246
-
Radiation dose of coronary CT angiography with a third-generation dual-source CT in a "real-world" patient population.Eur Radiol. 2019 Aug;29(8):4341-4348. doi: 10.1007/s00330-018-5856-6. Epub 2018 Nov 30. Eur Radiol. 2019. PMID: 30506216
-
Very low-dose coronary artery calcium scanning with high-pitch spiral acquisition mode: comparison between 120-kV and 100-kV tube voltage protocols.J Cardiovasc Comput Tomogr. 2013 Jan-Feb;7(1):32-8. doi: 10.1016/j.jcct.2012.11.004. Epub 2012 Dec 1. J Cardiovasc Comput Tomogr. 2013. PMID: 23333186
-
Coronary computed tomography angiography at 140 kV versus 120 kV: assessment of image quality and radiation exposure in overweight and moderately obese patients.Acta Radiol. 2014 Jun;55(5):554-62. doi: 10.1177/0284185113502745. Epub 2013 Sep 12. Acta Radiol. 2014. PMID: 24031049
Cited by
-
Low-Dose Radiation Advances in Coronary Computed Tomography Angiography in the Diagnosis of Coronary Artery Disease.Curr Cardiol Rev. 2019;15(4):304-315. doi: 10.2174/1573403X15666190222163737. Curr Cardiol Rev. 2019. PMID: 30806322 Free PMC article. Review.
-
Reduction of radiation dose using 80 kV tube voltage: a feasible strategy?Int J Cardiovasc Imaging. 2012 Feb;28(2):425-8. doi: 10.1007/s10554-011-9845-y. Epub 2011 Mar 19. Int J Cardiovasc Imaging. 2012. PMID: 21424152 Free PMC article. No abstract available.
-
Comparison of image quality between 70 kVp and 80 kVp: application to paediatric cardiac CT.Eur Radiol. 2014 Dec;24(12):3003-9. doi: 10.1007/s00330-014-3341-4. Epub 2014 Aug 31. Eur Radiol. 2014. PMID: 25173625
-
The research advancements on the advantages and limitations of coronary computed tomography angiography at low tube voltage: a literature review.Quant Imaging Med Surg. 2025 Jun 6;15(6):5859-5867. doi: 10.21037/qims-24-57. Epub 2025 May 26. Quant Imaging Med Surg. 2025. PMID: 40606334 Free PMC article. Review.
-
Optimizing contrast medium injection protocol individually with body weight for high-pitch prospective ECG-triggering coronary CT angiography.Int J Cardiovasc Imaging. 2013 Jun;29(5):1115-20. doi: 10.1007/s10554-012-0170-x. Epub 2013 Jan 4. Int J Cardiovasc Imaging. 2013. PMID: 23288418 Clinical Trial.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical