Rate responsive pacing using cardiac resynchronization therapy in patients with chronotropic incompetence and chronic heart failure
- PMID: 21551475
- DOI: 10.1093/europace/eur127
Rate responsive pacing using cardiac resynchronization therapy in patients with chronotropic incompetence and chronic heart failure
Abstract
Aims: Chronotropic incompetence (CI) is a common finding in patients with advanced chronic heart failure (CHF) and is associated with a worse functional capacity. Whether rate responsive pacing with cardiac resynchronization therapy (CRT) would acutely improve exercise performance in patients with advanced CHF and severe CI (<70% age-predicted maximum heart rate) is unknown.
Methods and results: Patients (n = 13) with CHF, a CRT device, and severe CI were randomized in a double-blind crossover pilot study to either DDD (control) or DDDR (rate responsive) pacing. Six minutes walk test (6MWT) distance, oxygen consumption at anaerobic threshold (VO(2) @ AT), and maximal oxygen consumption (VO(2) max) were measured. One week later, testing was repeated in the alternate pacing mode. Rate responsive pacing commenced with standard settings in only 9 of 13 (69%) patients. In these 9 subjects, 6MWT distance improved acutely from 358.5 ± 40.7 to 376.8 ± 24.5 m with DDDR pacing (P< 0.05). VO(2) max did not improve with DDDR pacing (14.0 ± 3.2 mL/kg/min) compared with DDD pacing (13.9 ± 3.0 mL/kg/min; P= 0.69). VO(2) @ AT tended towards improvement with DDDR pacing (10.8 ± 2.9 mL/kg/min) compared with DDD pacing (9.6 ± 1.8 mL/kg/min; P= 0.29). There was a linear relationship between the increase in heart rate at minute 3 during rate responsive pacing and improvement in VO(2) @ AT (r = 0.83, P< 0.05).
Conclusion: When rate responsive pacing using a CRT device is achieved in patients with advanced CHF and severe CI, parameters of aerobic exercise performance improve acutely. Routine exercise testing to ensure successful restoration of heart rate response may be beneficial to optimize CRT settings in this patient population.
Similar articles
-
[An intrapatient comparison of adaptation to aerobic and anaerobic exertion during 3 types of physiological cardiac stimulation in chronotropic failure of the sinus node: DDD, VVIR and DDDR].Cardiologia. 1997 Jan;42(1):51-7. Cardiologia. 1997. PMID: 9118155 Clinical Trial. Italian.
-
[DDD versus DDDR pacemaker stimulation: comparison of cardiopulmonary performance, incidence of atrial arrhythmias and quality of life].Z Kardiol. 1996 Apr;85(4):226-36. Z Kardiol. 1996. PMID: 8693765 Clinical Trial. German.
-
Impact of cardiac resynchronisation therapy on adaptation of circulatory and respiratory systems to exercise assessed by cardiopulmonary exercise test in patients with chronic heart failure.Kardiol Pol. 2008 Apr;66(4):406-12; discussion 413-4. Kardiol Pol. 2008. PMID: 18473269
-
Meta-analysis of effects of optimization of cardiac resynchronization therapy on left ventricular function, exercise capacity, and quality of life in patients with heart failure.Am J Cardiol. 2014 Mar 15;113(6):988-94. doi: 10.1016/j.amjcard.2013.12.006. Epub 2013 Dec 25. Am J Cardiol. 2014. PMID: 24461769 Review.
-
The importance of increased percentage of biventricular pacing to improve clinical outcomes in patients receiving cardiac resynchronization therapy.Curr Opin Cardiol. 2013 Jan;28(1):50-4. doi: 10.1097/HCO.0b013e32835b0b17. Curr Opin Cardiol. 2013. PMID: 23196776 Review.
Cited by
-
Chronotropic Incompetence During Exercise in Type 2 Diabetes: Aetiology, Assessment Methodology, Prognostic Impact and Therapy.Sports Med. 2015 Jul;45(7):985-95. doi: 10.1007/s40279-015-0328-5. Sports Med. 2015. PMID: 25834997 Review.
-
Atrioventricular Nodal Catheter Ablation in Atrial Fibrillation Complicating Congestive Heart Failure.J Atr Fibrillation. 2018 Jun 30;11(1):1813. doi: 10.4022/jafib.1813. eCollection 2018 Jun-Jul. J Atr Fibrillation. 2018. PMID: 30455833 Free PMC article.
-
Optimal Cardiac Resynchronization Therapy Pacing Rate in Non-Ischemic Heart Failure Patients: A Randomized Crossover Pilot Trial.PLoS One. 2015 Sep 18;10(9):e0138124. doi: 10.1371/journal.pone.0138124. eCollection 2015. PLoS One. 2015. PMID: 26382243 Free PMC article. Clinical Trial.
-
Chronotropic incompetence in Chagas disease: usefulness of dual sensor pacemaker based on volume minute and accelerometer.Rev Bras Cir Cardiovasc. 2015 Jul-Sep;30(3):III-VI. doi: 10.5935/1678-9741.20150011. Rev Bras Cir Cardiovasc. 2015. PMID: 26313733 Free PMC article. No abstract available.
-
Activated rate-response is associated with increased mortality risk in cardiac device carriers with acute heart failure.PLoS One. 2024 Apr 18;19(4):e0302321. doi: 10.1371/journal.pone.0302321. eCollection 2024. PLoS One. 2024. PMID: 38635729 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials