Comparative Study
. 2012 Jan;112(1):309-15.
doi: 10.1007/s00421-011-1983-3. Epub 2011 May 10.

Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based method

Affiliations
Comparative Study

Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based method

Heiko Gesche et al. Eur J Appl Physiol. 2012 Jan.

Abstract

Pulse transit time (PTT) and pulse wave velocity (PWV), respectively, were shown to have a correlation with systolic blood pressure (SBP) and have been reported to be suitable for indirect BP measurements. The aim of this study was to create a function between SBP and PWV, and to test its reliability for the determination of absolute SBP using a non-linear algorithm and a one-point calibration. 63 volunteers performed exercise to induce rises in BP. Arterial PTT was measured between the R-spike of the ECG and the plethysmographic curve of finger pulse-oximetry. The reference BP was measured using a cuff-based sphygmomanometric aneroid device. Data from 13 of the 63 volunteers served for the detection of the PWV-BP relationship. The created non-linear function was used to calculate BP values after individual correction for the BP offset in a group of 50 volunteers. Individual correlation coefficients for SBP measured by PTT (SBP(PTT)) and by cuff (SBP(CUFF)) varied between r = 0.69 and r = 0.99. Taking all data together, we found r = 0.83 (276 measurements in 50 volunteers). In the Bland-Altman plot, the limits of agreement were [Formula: see text]± 19.8 mmHg. In conclusion, comparing SBP values using the PTT-based method and those measured by cuff resulted in a significant correlation. However, the Bland-Altman plot shows relevant differences between both methods, which are partly due to greater variability of the SBP(PTT) measurement during intensified exercise. Results suggest that PTT can be used for measuring absolute SBP when performing an individual correction for the offset of the BP-PWV relation.

PubMed Disclaimer

References

    1. Hypertension. 2004 Jun;43(6):1239-45 - PubMed
    1. Dtsch Med Wochenschr. 2010 Dec;135(48):2406-12 - PubMed
    1. Int J Clin Monit Comput. 1997 Nov;14(4):251-4 - PubMed
    1. Ann Biomed Eng. 1988;16(4):341-7 - PubMed
    1. Med Biol Eng Comput. 2000 Sep;38(5):569-74 - PubMed

Publication types

LinkOut - more resources