Mechanical properties and tissue reinforcement of polypropylene grafts used for pelvic floor repair--an experimental study
- PMID: 21573997
- DOI: 10.1007/s10029-011-0829-6
Mechanical properties and tissue reinforcement of polypropylene grafts used for pelvic floor repair--an experimental study
Abstract
Purpose: Although grafts can be made from the same materials, the final weave of the product may play a role in the ultimate outcomes or complications. The aim of this experimental study is to evaluate and compare the mechanical properties of the meshes and their influence on biological characteristics as well.
Methods: Full-thickness abdominal wall defect with respect to the peritoneum was primary repaired in 144 male Wister rats using the overlay technique. Monofilament, multifilament and coated polypropylene grafts were used for the repairs. Graft shrinkage, thickness and tensile strength of the explants were analysed after 3 and 6 weeks.
Results: Monofilament grafts showed the highest shrinkage level of 11.97-18% during the investigated period. The thickness of monofilament explants increased gradually, with a maximum of 25.35% after 6 weeks. Multifilament and collagen explants did swell more rapidly in the third week, decreasing thereafter to 19.63-23.04% above the pre-implantation values after 6 weeks. The tensile strength showed a gradual increase over the course of the experiment in all groups.
Conclusion: After graft stabilisation, all samples had similar strength, regardless of the significant differences in the pre-implantation values of graft strength.
Similar articles
-
Tensile strength and host response towards different polypropylene implant materials used for augmentation of fascial repair in a rat model.Int Urogynecol J Pelvic Floor Dysfunct. 2007 Jun;18(6):619-26. doi: 10.1007/s00192-006-0202-y. Epub 2006 Oct 10. Int Urogynecol J Pelvic Floor Dysfunct. 2007. PMID: 17031489
-
Comparison of adhesive properties of five different prosthetic materials used in hernioplasty.J Invest Surg. 2005 Mar-Apr;18(2):89-95. doi: 10.1080/08941930590926357. J Invest Surg. 2005. PMID: 16036777
-
Histologic and biomechanical evaluation of a novel macroporous polytetrafluoroethylene knit mesh compared to lightweight and heavyweight polypropylene mesh in a porcine model of ventral incisional hernia repair.Hernia. 2011 Aug;15(4):423-31. doi: 10.1007/s10029-011-0787-z. Epub 2011 Jan 30. Hernia. 2011. PMID: 21279663 Free PMC article.
-
[Development of better tolerated prosthetic materials: applications in gynecological surgery].J Gynecol Obstet Biol Reprod (Paris). 2002 Oct;31(6):527-40. J Gynecol Obstet Biol Reprod (Paris). 2002. PMID: 12407323 Review. French.
-
Biology of biological meshes used in hernia repair.Surg Clin North Am. 2013 Oct;93(5):1211-5. doi: 10.1016/j.suc.2013.06.014. Epub 2013 Aug 12. Surg Clin North Am. 2013. PMID: 24035083 Review.
Cited by
-
Evaluation of three purely polypropylene meshes of different pore sizes in an onlay position in a New Zealand white rabbit model.Hernia. 2014;18(6):855-64. doi: 10.1007/s10029-014-1278-9. Epub 2014 Jul 18. Hernia. 2014. PMID: 25033941
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources