. 2011 Sep;41(9):1174-82.
doi: 10.1007/s00247-011-2063-x. Epub 2011 May 19.

Reducing abdominal CT radiation dose with the adaptive statistical iterative reconstruction technique in children: a feasibility study

Affiliations

Reducing abdominal CT radiation dose with the adaptive statistical iterative reconstruction technique in children: a feasibility study

Gregory A Vorona et al. Pediatr Radiol. 2011 Sep.

Abstract

Background: The use of the adaptive statistical iterative reconstruction (ASIR) algorithm has been shown to reduce radiation doses in adults undergoing abdominal CT studies while preserving image quality. To our knowledge, no studies have been done to validate the use of ASIR in children.

Objective: To retrospectively evaluate differences in radiation dose and image quality in pediatric CT abdominal studies utilizing 40% ASIR compared with filtered-back projection (FBP).

Materials and methods: Eleven patients (mean age 8.5 years, range 2-17 years) had separate 40% ASIR and FBP enhanced abdominal CT studies on different days between July 2009 and October 2010. The ASIR studies utilized a 38% mA reduction in addition to our pediatric protocol mAs. Study volume CT dose indexes (CTDI(vol)) and dose-length products (DLP) were recorded. A consistent representative image was obtained from each study. The images were independently evaluated by two radiologists in a blinded manner for diagnostic utility, image sharpness and image noise.

Results: The average CTDI(vol) and DLP for the 40% ASIR studies were 4.25 mGy and 185.04 mGy-cm, compared with 6.75 mGy and 275.79 mGy-cm for the FBP studies, representing 37% and 33% reductions in both, respectively. The radiologists' assessments of subjective image quality did not demonstrate any significant differences between the ASIR and FBP images.

Conclusion: In our experience, the use of 40% ASIR with a 38% decrease in mA lowers the radiation dose for children undergoing enhanced abdominal examinations by an average of 33%, while maintaining diagnostically acceptable images.

PubMed Disclaimer

References

    1. Radiology. 2010 Nov;257(2):373-83 - PubMed
    1. AJR Am J Roentgenol. 2010 Jan;194(1):191-9 - PubMed
    1. AJR Am J Roentgenol. 2010 Sep;195(3):713-9 - PubMed
    1. AJR Am J Roentgenol. 2009 Sep;193(3):764-71 - PubMed