[Qualitative control over quantitative MR tomography: in vitro and in vivo checks on relaxation time measurements]
- PMID: 2163072
- DOI: 10.1055/s-2008-1046946
[Qualitative control over quantitative MR tomography: in vitro and in vivo checks on relaxation time measurements]
Abstract
In-vivo and in-vitro checks on measurements of relaxation times in MRI have been carried out as part of quality control. For the in-vitro experiments we used 13 reagents consisting of varying concentrations of copper sulphate manganese sulphate, gadolinium solutions and various oils as well as water. For in-vivo measurements we used liver parenchyma and fat in three subjects. Eight different sequences were performed; intra- and interindividual variations and comparison of three similar MRI units showed average variations of 2-3% for T1 and 3-4% variations for T2 relaxation times. This margin of error makes quantitative MRI acceptable for clinical use.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
