High complication rate after biplanar open wedge high tibial osteotomy stabilized with a new spacer plate (Position HTO plate) without bone substitute
- PMID: 21663721
- DOI: 10.1016/j.arthro.2011.01.008
High complication rate after biplanar open wedge high tibial osteotomy stabilized with a new spacer plate (Position HTO plate) without bone substitute
Abstract
Purpose: We performed a prospective clinical and radiographic evaluation after open wedge high tibial osteotomy (HTO) using the new Position HTO plate (Aesculap, Tuttlingen, Germany) without bone transplantation.
Methods: Thirty-five open wedge HTOs with the Position HTO plate were performed without bone wedges. The mean patient age was 44.6 ± 9.2 years at the time of osteotomy, which was planned with mediCAD II software (Hectec, Niederviehbach, Germany). The Hospital for Special Surgery score, Lysholm-Gillquist score, Tegner activity level, and International Knee Documentation Committee subjective score were used for clinical assessment. We evaluated radiographs obtained preoperatively and at 2, 6, and 12 months postoperatively using full-weight-bearing anteroposterior whole-leg views and anteroposterior and lateral views of the knee. For statistical analyses, JMP 8.0.1 (SAS, Cary, NC) was used.
Results: We observed an overall complication rate of 34% and a plate-related complication rate of 23%. Plate-related complications included loss of correction, fracture of the tibial plateau, screw failure, malunion, and fracture of the lateral cortical bone. A significant difference in the mechanical tibiofemoral angle of -1.3° ± 1.4° (P < .001) was found between the follow-up at 2 and 6 months. The mean Hospital for Special Surgery score was 74.8 ± 11.7 preoperatively, and it increased to 87.8 ± 11.0 (P < .001). The mean score on the Lysholm-Gillquist knee functional scoring scale was 55.5 ± 21.7 preoperatively, and it improved to 73.0 ± 23.9 (P < .001). The Tegner activity level was 2.6 ± 0.9 preoperatively, and it improved significantly at final follow-up to 3.7 ± 1.8 (P < .02). The International Knee Documentation Committee subjective score was 43.0 ± 14.9 preoperatively, and it increased to 66.1 ± 21 (P < .001).
Conclusions: We have shown a high plate-related complication rate and a significant loss of correction between 2 and 6 months of follow-up after open wedge HTO using the new Position HTO plate without bone wedges. The preoperatively planned mechanical tibiofemoral angle was not achieved. Despite these complications, the clinical outcome improved significantly. The Position HTO plate cannot be recommended with the presented technique.
Level of evidence: Level IV, therapeutic case series.
Copyright © 2011 Arthroscopy Association of North America. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Comparative study of medial opening-wedge high tibial osteotomy using 2 different implants.Arthroscopy. 2013 Jun;29(6):1063-71. doi: 10.1016/j.arthro.2013.02.020. Epub 2013 Apr 26. Arthroscopy. 2013. PMID: 23623294
-
Open-wedge osteotomy using an internal plate fixator in patients with medial-compartment gonarthritis and varus malalignment: 3-year results with regard to preoperative arthroscopic and radiographic findings.Arthroscopy. 2010 Dec;26(12):1607-16. doi: 10.1016/j.arthro.2010.05.006. Arthroscopy. 2010. PMID: 20926232 Clinical Trial.
-
High tibial open wedge osteotomy below the tibial tubercle: clinical and radiographic results.Knee Surg Sports Traumatol Arthrosc. 2013 Jan;21(1):57-63. doi: 10.1007/s00167-011-1453-9. Epub 2011 Mar 8. Knee Surg Sports Traumatol Arthrosc. 2013. PMID: 21384234
-
Accuracy of frontal and sagittal plane correction in open-wedge high tibial osteotomy.Arthroscopy. 2004 Apr;20(4):366-72. doi: 10.1016/j.arthro.2004.01.024. Arthroscopy. 2004. PMID: 15067275 Review.
-
Comparison of navigated and conventional high tibial osteotomy for the treatment of osteoarthritic knees with varus deformity: A meta-analysis.Int J Surg. 2018 Jul;55:211-219. doi: 10.1016/j.ijsu.2018.03.024. Epub 2018 Mar 16. Int J Surg. 2018. PMID: 29555521 Review.
Cited by
-
Lateral hinge fracture in medial opening wedge high tibial osteotomy: a narrative review.EFORT Open Rev. 2023 Jul 3;8(7):572-580. doi: 10.1530/EOR-22-0103. EFORT Open Rev. 2023. PMID: 37395709 Free PMC article. Review.
-
Biplanar Open Wedge High Tibial Osteotomy in the Medial Compartment Osteoarthritis of the Knee Joint: Comparison between the Aescula and TomoFix Plate.Clin Orthop Surg. 2015 Jun;7(2):185-90. doi: 10.4055/cios.2015.7.2.185. Epub 2015 May 18. Clin Orthop Surg. 2015. PMID: 26217464 Free PMC article.
-
Disparity between Preoperative Target Correction Amount and Postoperative Correction Amount in Open Wedge High Tibial Osteotomy.Knee Surg Relat Res. 2019 Jun 1;31(2):126-131. doi: 10.5792/ksrr.18.034. Knee Surg Relat Res. 2019. PMID: 30893991 Free PMC article.
-
Design optimization of high tibial osteotomy plates using finite element analysis for improved biomechanical effect.J Orthop Surg Res. 2019 Jul 16;14(1):219. doi: 10.1186/s13018-019-1269-8. J Orthop Surg Res. 2019. PMID: 31311570 Free PMC article.
-
Biomechanical properties of a new anatomical locking metal block plate for opening wedge high tibial osteotomy: uniplane osteotomy.Knee Surg Relat Res. 2014 Sep;26(3):155-61. doi: 10.5792/ksrr.2014.26.3.155. Epub 2014 Aug 29. Knee Surg Relat Res. 2014. PMID: 25229045 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials