Corneal biomechanical evaluation in healthy thin corneas compared with matched keratoconus cases
- PMID: 21670900
- DOI: 10.1590/s0004-27492011000100003
Corneal biomechanical evaluation in healthy thin corneas compared with matched keratoconus cases
Abstract
Purpose: To evaluate and compare corneal hysteresis (CH) and corneal resistance factor (CRF) in healthy eyes with a central corneal thickness (CCT) < 505 µm with CH and CRF in gender-, age-, and CCT-matched keratoconus cases, and to estimate the sensitivity and specificity of these parameters for discriminating between the two groups.
Methods: Prospective, comparative case series. In total 46 eyes from 30 healthy patients with CCT < 505 µm, and 42 eyes from 30 CCT-, gender- and age-matched keratoconus cases were enrolled. Biomechanical metrics (CH and CRF) were measured using the Ocular Response Analyzer (ORA) and then compared. A receiver operating characteristic (ROC) curve was used to identify cut-off points to maximize the sensitivity and specificity for discriminating between the groups.
Results: The CCT was 485.96 ± 17.61 µm (range, 438 - 505) in healthy thin corneas and 483.64 ± 16.19 µm (range, 452 - 505) in keratoconus; p=0.5225. CH was 8.63 ± 1.23 mmHg (range, 5.95 - 12.2) and 8.07 ± 1.17 mmHg (range, 4.9 - 9.85), respectively; p=0.0312. CRF was 8.43 ± 1.29 mmHg (range, 5.45 - 11.10) and 7.22 ± 1.34 mmHg (range, 4.7 - 9.45), respectively; p<0.001. ROC curve analysis showed a poor overall predictive accuracy of CH (cut-off, 8.95 mmHg; sensitivity, 63%; specificity, 23.8%; test accuracy, 44.30%) and CRF (cut-off, 7.4 mmHg; sensitivity, 28.3%; specificity, 40.5%; test accuracy, 34.12%) for detecting keratoconus in the eyes studied.
Conclusion: CH and CRF were statistically lower in keratoconus than in healthy thin corneas. However, CH and CRF offered very low sensitivity and specificity for discriminating the groups.
Similar articles
-
Corneal biomechanical metrics and anterior segment parameters in mild keratoconus.Ophthalmology. 2010 Apr;117(4):673-9. doi: 10.1016/j.ophtha.2009.09.023. Ophthalmology. 2010. PMID: 20138369
-
Ability of corneal biomechanical metrics and anterior segment data in the differentiation of keratoconus and healthy corneas.Arq Bras Oftalmol. 2010 Jul-Aug;73(4):333-7. doi: 10.1590/s0004-27492010000400006. Arq Bras Oftalmol. 2010. PMID: 20944935
-
Improved keratoconus detection by ocular response analyzer testing after consideration of corneal thickness as a confounding factor.J Refract Surg. 2012 Mar;28(3):202-8. doi: 10.3928/1081597X-20120103-03. Epub 2012 Jan 9. J Refract Surg. 2012. PMID: 22230059
-
Identification of biomechanical properties of the cornea: the ocular response analyzer.Curr Eye Res. 2012 Jul;37(7):553-62. doi: 10.3109/02713683.2012.669007. Epub 2012 May 4. Curr Eye Res. 2012. PMID: 22559332 Review.
-
Brillouin microscopy: assessing ocular tissue biomechanics.Curr Opin Ophthalmol. 2018 Jul;29(4):299-305. doi: 10.1097/ICU.0000000000000489. Curr Opin Ophthalmol. 2018. PMID: 29771749 Free PMC article. Review.
Cited by
-
A Novel Approach to Fabricate Early Keratoconus Phantom Models.Transl Vis Sci Technol. 2025 Apr 1;14(4):18. doi: 10.1167/tvst.14.4.18. Transl Vis Sci Technol. 2025. PMID: 40244577 Free PMC article.
-
Impaired corneal biomechanical properties and the prevalence of keratoconus in mitral valve prolapse.J Ophthalmol. 2014;2014:402193. doi: 10.1155/2014/402193. Epub 2014 Apr 17. J Ophthalmol. 2014. PMID: 24864193 Free PMC article.
-
Advances in Biomechanical Parameters for Screening of Refractive Surgery Candidates: A Review of the Literature, Part III.Med Hypothesis Discov Innov Ophthalmol. 2019 Fall;8(3):219-240. Med Hypothesis Discov Innov Ophthalmol. 2019. PMID: 31598522 Free PMC article. Review.
-
Novel dynamic corneal response parameters in a practice use: a critical review.Biomed Eng Online. 2019 Feb 13;18(1):17. doi: 10.1186/s12938-019-0636-3. Biomed Eng Online. 2019. PMID: 30760270 Free PMC article. Review.
-
In vivo measurement of regional corneal tangent modulus.Sci Rep. 2017 Nov 2;7(1):14974. doi: 10.1038/s41598-017-14750-w. Sci Rep. 2017. PMID: 29097675 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources