Multicenter Study
. 2011 Jun 15;7(3):301-6.
doi: 10.5664/JCSM.1078.

Sleep staging based on autonomic signals: a multi-center validation study

Affiliations
Multicenter Study

Sleep staging based on autonomic signals: a multi-center validation study

Jan Hedner et al. J Clin Sleep Med. .

Abstract

Study objectives: One of the most important caveats of ambulatory devices is the inability to record and stage sleep. We assessed an algorithm determining 4 different stages: wake, light sleep, deep sleep, and REM sleep using signals derived from the portable monitor Watch-PAT100 (PAT recorder).

Methods: Participants (38 normal subjects and 189 patients with obstructive sleep apnea [OSA]) underwent simultaneous overnight recordings with polysomnography (PSG) and the PAT recorder in a study originally designed to assess the accuracy of the PAT recorder in diagnosing OSA. Light/deep sleep and REM sleep from the PAT recorder recording were automatically scored based on features extracted from time series of peripheral arterial tone amplitudes and inter pulse periods. The PSG scored sleep stages 1 and 2 were classified as light sleep for epoch-by-epoch comparisons.

Results: The overall agreement in detecting light/deep and REM sleep were 88.6% ± 5.9% and 88.7% ± 5.5%, respectively. There was a good agreement between PSG and the PAT recorder in quantifying sleep efficiency (78.4% ± 9.9% vs. 78.8% ± 13.4%), REM latency (237 ± 148 vs. 225 ± 159 epochs), and REM percentage (14.4% ± 6.5% vs. 19.3% ± 8.7%). OSA severity did not affect the sensitivity and specificity of the algorithm. The Cohen κ coefficient for detecting all sleep stages: sleep from wake, REM from NREM sleep, and deep from light sleep were 0.48, 0.55, 0.59, and 0.46, respectively.

Conclusions: Analysis of autonomic signals from the PAT recorder can detect sleep stages with moderate agreement to more standard techniques in normal subjects and OSA patients. This novel algorithm may provide insights on sleep and sleep architecture when applying the PAT recorder for OSA diagnosis.

Keywords: Ambulatory recording; REM sleep; autonomic; deep sleep stages; diagnosis; light sleep stages; peripheral arterial tone; sleep apnea.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Interclass correlation and Bland Altman plot for the scoring of respiratory disturbance index (RDI) The interclass correlation between the methods was r = 0.87, p < 0.01. On the right, the X axis represents the mean RDI between the PSG and the PAT recorder, and the Y axis the difference between methods.

References

    1. Loredo JS, Nelesen R, Ancoli-Israel S, et al. Sleep quality and blood pressure dipping in normal adults. Sleep. 2004;27:1097–103. - PubMed
    1. Cartwright R, Baehr E, Kirkby J, et al. REM sleep reduction, mood regulation and remission in untreated depression. Psychiatry Res. 2003;121:159–67. - PubMed
    1. McArdle N, Douglas NJ. Effect of continuous positive airway pressure on sleep architecture in the sleep apnea-hypopnea syndrome: a randomized controlled trial. Am J Respir Crit Care Med. 2001;164:1459–63. - PubMed
    1. Collop NA, Anderson WM, Boehlecke B, et al. Clinical guidelines for the use of unattended portable monitors in the diagnosis of obstructive sleep apnea in adult patients. Portable Monitoring Task Force of the American Academy of Sleep Medicine. J Clin Sleep Med. 2007;3:737–47. - PMC - PubMed
    1. Bar A, Pillar G, Dvir I, et al. Evaluation of a portable device based on peripheral arterial tone for unattended home sleep studies. Chest. 2003;123:695–703. - PubMed

Publication types