Comparison of closure strategies after balloon aortic valvuloplasty: suture mediated versus collagen based versus manual
- PMID: 21681898
- DOI: 10.1002/ccd.22940
Comparison of closure strategies after balloon aortic valvuloplasty: suture mediated versus collagen based versus manual
Abstract
Objectives: To compare complication rates of a collagen-mediated closure device, suture-mediated closure technique and manual compression for access site management following balloon aortic valvuloplasty (BAV).
Background: Vascular complications, including perforation, limb ischemia, ateriovenous fistulas, and pseudoaneurysm, are a common source of morbidity in patients undergoing BAV with large bore femoral artery access. The available options for closure of the access site include manual compression, suture-mediated closure devices, and recently reported collagen-based closure devices.
Methods: The study cohort consisted of 333 patients with severe aortic stenosis undergoing BAV. Patients were divided into four groups according to access closure method: (1) Manual compression in 64 (19.2%); (2) Proglide 6 Fr in 162 (48.6%); (3) Angio-Seal 8 Fr in 89 (26.75%); and (4) Prostar 10 Fr in 18 (5.4%).
Results: There were no significant differences in baseline characteristics among the groups; although the Prostar group was had more males and the Angio-Seal group reported a lower incidence of renal failure and peripheral vascular disease. The Angio-Seal group had shorter procedural duration times compared with the other groups. In the Angio-Seal group (n = 47, 52.8%), a 10 Fr sheath size was primarily used to access the femoral artery, as compared with the manual compression and Proglide groups, which used a 12 Fr sheath in 29 (45.3%) and 66 patients (40.7%), respectively. A 13 Fr sheath was used in the Prostar group (n = 11, 61.1%), P < 0.001. The device failure rate was significantly higher in the Proglide group (n = 20, 12.3%) as compared with Angio-Seal (n = 3, 3.3%) and Prostar groups (n = 1, 5.5%), P = 0.03. Serious vascular complications requiring intervention were significantly higher in the manual compression group (n = 11, 17.1%) compared with the Proglide (n = 11, 6.7%), Angio-Seal (n = 5, 5.6%), and Prostar groups (n = 5, 1%), P < 0.001. The requirement for blood transfusion in the Angio-Seal group was significantly lower (n = 7, 7.8%) as compared with the manual compression (n = 20, 31.2%), Proglide (n = 35, 21.6%), and Prostar groups (n = 5, 27.7%), P < 0.001.
Conclusion: In our case series, collagen-based closure devices and suture-mediated closure devices had fewer vascular complications than manual compression for hemostasis following BAV. Patient selection for collagen-based versus suture-mediated closure requires further study.
Copyright © 2011 Wiley-Liss, Inc.
Comment in
-
Closing in on large hole closure.Catheter Cardiovasc Interv. 2011 Jul 1;78(1):125-6. doi: 10.1002/ccd.23251. Catheter Cardiovasc Interv. 2011. PMID: 21681899 No abstract available.
Similar articles
-
"Double wire" angio-seal closure technique after balloon aortic valvuloplasty.Catheter Cardiovasc Interv. 2010 Mar 1;75(4):488-92. doi: 10.1002/ccd.22295. Catheter Cardiovasc Interv. 2010. PMID: 19937771
-
Large calibre arterial access device closure for percutaneous aortic valve interventions: use of the Prostar system in 118 cases.Catheter Cardiovasc Interv. 2012 Jan 1;79(1):143-9. doi: 10.1002/ccd.23105. Epub 2011 Dec 8. Catheter Cardiovasc Interv. 2012. PMID: 21538787
-
Vascular complications after balloon aortic valvuloplasty in recent years: Incidence and comparison of two hemostatic devices.Catheter Cardiovasc Interv. 2018 May 1;91(6):E49-E55. doi: 10.1002/ccd.27328. Epub 2017 Oct 5. Catheter Cardiovasc Interv. 2018. PMID: 28980387
-
Combination of Suture-Plug Vascular Closure Devices Versus Suture-Only Devices Following Transcatheter Aortic Valve Implantation: A Systematic Review and Meta-Analysis.Catheter Cardiovasc Interv. 2025 Aug;106(2):1088-1097. doi: 10.1002/ccd.31655. Epub 2025 Jun 2. Catheter Cardiovasc Interv. 2025. PMID: 40457608 Review.
-
Vascular Closure Devices for Large Bore Femoral Access After Transfemoral Transcatheter Aortic Valve Replacement: A Network Meta-Analysis.Catheter Cardiovasc Interv. 2025 Jul;106(1):423-432. doi: 10.1002/ccd.31565. Epub 2025 Apr 27. Catheter Cardiovasc Interv. 2025. PMID: 40289515
Cited by
-
The use of vascular closure devices and impact on major bleeding and net adverse clinical events (NACEs) in balloon aortic valvuloplasty: a sub-analysis of the BRAVO study.Catheter Cardiovasc Interv. 2014 Jan 1;83(1):148-53. doi: 10.1002/ccd.24892. Epub 2013 Mar 25. Catheter Cardiovasc Interv. 2014. PMID: 23436434 Free PMC article.
-
Vascular Closure Devices versus Manual Compression in Cardiac Interventional Procedures: Systematic Review and Meta-Analysis.Cardiovasc Ther. 2022 Sep 9;2022:8569188. doi: 10.1155/2022/8569188. eCollection 2022. Cardiovasc Ther. 2022. PMID: 36134143 Free PMC article.
-
Percutaneous balloon aortic valvuloplasty in the era of transcatheter aortic valve implantation: a narrative review.Open Heart. 2016 Dec 7;3(2):e000421. doi: 10.1136/openhrt-2016-000421. eCollection 2016. Open Heart. 2016. PMID: 28008354 Free PMC article.
-
Safety and Efficacy of the Prostar XL Vascular Closing Device for Percutaneous Closure of Large Arterial Access Sites.Radiol Res Pract. 2013;2013:875484. doi: 10.1155/2013/875484. Epub 2013 Jan 14. Radiol Res Pract. 2013. PMID: 23401769 Free PMC article.
-
Adjunctive use of Angio-Seal closure device following transcatheter aortic valve implantation via percutaneous transfemoral approach with incomplete hemostasis after modified Perclose ProGlide preclosure technique.Proc (Bayl Univ Med Cent). 2019 Jan 11;32(1):34-36. doi: 10.1080/08998280.2018.1524102. eCollection 2019 Jan. Proc (Bayl Univ Med Cent). 2019. PMID: 30956577 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical