The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems
- PMID: 2168451
- DOI: 10.1097/00004872-199007000-00004
The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems
Abstract
With the increasing manufacture of expensive systems for the measurement of ambulatory blood pressure there is a need for potential purchasers to be able to satisfy themselves that the systems have been evaluated according to agreed criteria. The British Hypertension Society has, therefore, drawn up a protocol of requirements for the evaluation of these devices. This protocol incorporates many features of the American National Standard for Non-Automated Sphygmomanometers but includes many additional features, such as strict criteria for observer training, interdevice variability testing before and after a month of ambulatory use, and a new system of analysis which permits the test system to be graded. It is recommended that manufacturers of ambulatory blood pressure measuring devices should obtain an unbiased evaluation according to a recognized standard before a device is marketed.
Comment in
-
British hypertension protocol.J Hypertens. 1991 Jun;9(6):575-7. doi: 10.1097/00004872-199106000-00017. J Hypertens. 1991. PMID: 1653300 No abstract available.
-
The assessment of portable, non-invasive blood pressure recorders.J Hypertens. 1990 Jul;8(7):591-3. doi: 10.1097/00004872-199007000-00001. J Hypertens. 1990. PMID: 2168448 No abstract available.
Similar articles
-
Validation requirements for ambulatory blood pressure measuring systems.J Hypertens Suppl. 1991 Dec;9(8):S13-5. J Hypertens Suppl. 1991. PMID: 1795192
-
Evaluation of blood pressure measuring devices with special reference to ambulatory systems.J Hypertens Suppl. 1990 Dec;8(7):S133-9. J Hypertens Suppl. 1990. PMID: 2095381
-
Bench and ambulatory field evaluation of the A & D TM-2420 automated sphygmomanometer.J Hypertens. 1990 Jul;8(7):599-605. doi: 10.1097/00004872-199007000-00003. J Hypertens. 1990. PMID: 2168450
-
Evaluation of blood pressure measuring devices.Clin Exp Hypertens. 1993 Nov;15(6):1087-97. doi: 10.3109/10641969309037096. Clin Exp Hypertens. 1993. PMID: 8268876 Review.
-
Factors influencing validation of ambulatory blood pressure measuring devices.J Hypertens. 1995 Nov;13(11):1235-40. J Hypertens. 1995. PMID: 8984119 Review.
Cited by
-
Continuous blood pressure prediction system using Conv-LSTM network on hybrid latent features of photoplethysmogram (PPG) and electrocardiogram (ECG) signals.Sci Rep. 2024 Jul 16;14(1):16450. doi: 10.1038/s41598-024-66514-y. Sci Rep. 2024. PMID: 39014018 Free PMC article.
-
STRIDE BP international initiative for accurate blood pressure measurement: Systematic review of published validation studies of blood pressure measuring devices.J Clin Hypertens (Greenwich). 2019 Nov;21(11):1616-1622. doi: 10.1111/jch.13710. Epub 2019 Oct 26. J Clin Hypertens (Greenwich). 2019. PMID: 31654494 Free PMC article.
-
Short-term blood pressure changes have a more strong impact on stroke and its subtypes than long-term blood pressure changes.Clin Cardiol. 2019 Oct;42(10):925-933. doi: 10.1002/clc.23242. Epub 2019 Jul 30. Clin Cardiol. 2019. PMID: 31361034 Free PMC article.
-
Validation of automated oscillometric sphygmomanometer (HDBPM) for arterial pressure measurement during haemodialysis.Med Biol Eng Comput. 2000 Jan;38(1):98-101. doi: 10.1007/BF02344696. Med Biol Eng Comput. 2000. PMID: 10829398 Clinical Trial.
-
Novel methods of testing and calibration of oscillometric blood pressure monitors.PLoS One. 2018 Aug 6;13(8):e0201123. doi: 10.1371/journal.pone.0201123. eCollection 2018. PLoS One. 2018. PMID: 30080862 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical