Evaluation of a novel needle guide for ultrasound-guided phantom vessel cannulation
- PMID: 21692759
- DOI: 10.1111/j.1365-2044.2011.06781.x
Evaluation of a novel needle guide for ultrasound-guided phantom vessel cannulation
Abstract
We evaluated a novel, sled-mounted needle guide for ultrasound-guided vessel cannulation. Fifty medical students were randomly assigned to use ultrasound with the sled (sled group, n = 23) or ultrasound without the sled (control group, n = 27) for vessel cannulation in a phantom. For each of 15 attempts we recorded cannulation time and designated a successful cannulation as 1 and a failure as 0. Our primary outcome was the mean overall success rate. The median (IQR [range]) number of successes in the sled group and control group were 15.0 (13.0-15.0 [11.0-15.0]) and 11.0 (9.0-13.0 [6.0-15.0]), respectively (p < 0.001). Cannulation time decreased from the first to the last attempt in the sled group from 7.0 s (6.0-10.0 [4.0-16]) s to 4.0 s (3.0-4.0 [1.0-6.0]) s and in the control group from 35.0 s (27.0-35.0 [11.0-35.0]) s to 7.0 s (5.0-10.0 [3.0-25.0]) s. The sled group demonstrated a shorter cannulation time at each attempt (p < 0.001). The novel sled improved the success rate and efficiency of ultrasound-guided phantom vessel cannulation.
© 2011 The Authors. Anaesthesia © 2011 The Association of Anaesthetists of Great Britain and Ireland.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources