. 2011 Jul;37(7):1189-98.
doi: 10.1016/j.jcrs.2011.04.022.

Femtosecond laser capsulotomy

Affiliations

Femtosecond laser capsulotomy

Neil J Friedman et al. J Cataract Refract Surg. 2011 Jul.

Erratum in

  • J Cataract Refract Surg. 2011 Sep;37(9):1742

Abstract

Purpose: To evaluate a femtosecond laser system to create the capsulotomy.

Setting: Porcine and cadaver eye studies were performed at OptiMedica Corp., Santa Clara, California, USA; the human trial was performed at the Centro Laser, Santo Domingo, Dominican Republic.

Design: Experimental and clinical study.

Methods: Capsulotomies performed by an optical coherence tomography-guided femtosecond laser were evaluated in porcine and human cadaver eyes. Subsequently, the procedure was performed in 39 patients as part of a prospective randomized study of femtosecond laser-assisted cataract surgery. The accuracy of the capsulotomy size, shape, and centration were quantified and capsulotomy strength was assessed in the porcine eyes.

Results: Laser-created capsulotomies were significantly more precise in size and shape than manually created capsulorhexes. In the patient eyes, the deviation from the intended diameter of the resected capsule disk was 29 μm ± 26 (SD) for the laser technique and 337 ± 258 μm for the manual technique. The mean deviation from circularity was 6% and 20%, respectively. The center of the laser capsulotomies was within 77 ± 47 μm of the intended position. All capsulotomies were complete, with no radial nicks or tears. The strength of laser capsulotomies (porcine subgroup) decreased with increasing pulse energy: 152 ± 21 mN for 3 μJ, 121 ± 16 mN for 6 μJ, and 113 ± 23 mN for 10 μJ. The strength of the manual capsulorhexes was 65 ± 21 mN.

Conclusion: The femtosecond laser produced capsulotomies that were more precise, accurate, reproducible, and stronger than those created with the conventional manual technique.

PubMed Disclaimer

Comment in

  • Femtosecond laser capsulotomy.
    Bala C. Bala C. J Cataract Refract Surg. 2021 Aug 1;47(8):979-981. doi: 10.1097/j.jcrs.0000000000000728. J Cataract Refract Surg. 2021. PMID: 34290194 No abstract available.

Publication types