. 2013 Feb;22(2):127-32.
doi: 10.1097/IJG.0b013e3182254ba3.

Repeatability and reproducibility of Goldmann applanation, dynamic contour, and ocular response analyzer tonometry

Affiliations

Repeatability and reproducibility of Goldmann applanation, dynamic contour, and ocular response analyzer tonometry

Allen Shawlun Wang et al. J Glaucoma. 2013 Feb.

Abstract

Purpose: To evaluate the repeatability and interoperator reproducibility of the Pascal dynamic contour tonometry (DCT), ocular response analyzer (ORA), and Goldmann applanation tonometer (GAT) in a single population of normal individuals.

Methods: The study included 52 eyes from 26 normal individuals. One operator measured the intraocular pressure (IOP) with each tonometer 3 times while 2 additional operators each measured the IOP with each tonometer once. Repeatability and reproducibility were assessed by the coefficient of variation (CV) and intraclass correlation coefficient (ICC). Agreement among tonometers was also assessed using Bland-Altman plots.

Results: The mean age of included participants was 31.5±8.8 years and 15 (58%) were female individuals. In general, both intraoperator repeatability and interoperator reproducibility were significantly higher for DCT compared with the other tonometers. Intraoperator DCT (CV=3.7, ICC=0.89), GAT (CV=9.7, ICC=0.79), IOPg (CV=7.0, ICC=0.79), and IOPcc (CV=9.8, ICC=0.57). Interoperator DCT (CV=6.1, ICC=0.73), GAT (CV=9.0, ICC=0.82), and IOPg (CV=10.8, ICC=0.63), IOPcc (CV=11.7, ICC=0.49).

Conclusion: Overall, DCT was significantly more repeatable and reproducible than GAT, IOPg and IOPcc. The better reproducibility of the DCT may result in more precise measurements for monitoring IOP changes over time compared to GAT and ORA.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Bland-Altman plots of the Intraocular Pressure (IOP) measurements by the Dynamic Contour Tonometer (DCT), Goldmann Applanation Tonometer (GAT), Ocular Response Analyzer - Goldman Estimated (IOPg) and the Ocular Response Analyzer – cornea corrected (IOPcc) for the Intra-Operator group of measurements.
Figure 1
Figure 1
Bland-Altman plots of the Intraocular Pressure (IOP) measurements by the Dynamic Contour Tonometer (DCT), Goldmann Applanation Tonometer (GAT), Ocular Response Analyzer - Goldman Estimated (IOPg) and the Ocular Response Analyzer – cornea corrected (IOPcc) for the Intra-Operator group of measurements.
Figure 1
Figure 1
Bland-Altman plots of the Intraocular Pressure (IOP) measurements by the Dynamic Contour Tonometer (DCT), Goldmann Applanation Tonometer (GAT), Ocular Response Analyzer - Goldman Estimated (IOPg) and the Ocular Response Analyzer – cornea corrected (IOPcc) for the Intra-Operator group of measurements.
Figure 1
Figure 1
Bland-Altman plots of the Intraocular Pressure (IOP) measurements by the Dynamic Contour Tonometer (DCT), Goldmann Applanation Tonometer (GAT), Ocular Response Analyzer - Goldman Estimated (IOPg) and the Ocular Response Analyzer – cornea corrected (IOPcc) for the Intra-Operator group of measurements.
Figure 2
Figure 2
Bland-Altman plots of the Intraocular Pressure (IOP) measurements by the Dynamic Contour Tonometer (DCT), Goldmann Applanation Tonometer (GAT), Ocular Response Analyzer - Goldman Estimated (IOPg) and the Ocular Response Analyzer – cornea corrected (IOPcc) for the Inter-Operator group of measurements.
Figure 2
Figure 2
Bland-Altman plots of the Intraocular Pressure (IOP) measurements by the Dynamic Contour Tonometer (DCT), Goldmann Applanation Tonometer (GAT), Ocular Response Analyzer - Goldman Estimated (IOPg) and the Ocular Response Analyzer – cornea corrected (IOPcc) for the Inter-Operator group of measurements.
Figure 2
Figure 2
Bland-Altman plots of the Intraocular Pressure (IOP) measurements by the Dynamic Contour Tonometer (DCT), Goldmann Applanation Tonometer (GAT), Ocular Response Analyzer - Goldman Estimated (IOPg) and the Ocular Response Analyzer – cornea corrected (IOPcc) for the Inter-Operator group of measurements.
Figure 2
Figure 2
Bland-Altman plots of the Intraocular Pressure (IOP) measurements by the Dynamic Contour Tonometer (DCT), Goldmann Applanation Tonometer (GAT), Ocular Response Analyzer - Goldman Estimated (IOPg) and the Ocular Response Analyzer – cornea corrected (IOPcc) for the Inter-Operator group of measurements.

References

    1. Goldmann H, Schmidt T. Applanation tonometry. Ophthalmologica. 1957;134(4):221–242. - PubMed
    1. Kotecha A, Elsheikh A, Roberts CR, et al. Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer. Invest Ophthalmol Vis Sci. 2006;47(12):5337–5347. - PubMed
    1. Kaufmann C, Bachmann LM, Thiel MA. Comparison of dynamic contour tonometry with goldmann applanation tonometry. Invest Ophthalmol Vis Sci. 2004;45(9):3118–3121. - PubMed
    1. Kotecha A, White ET, Shewry JM, Garway-Heath DF. The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry. Br J Ophthalmol. 2005;89(12):1572–1575. - PMC - PubMed
    1. Johannesson G, Hallberg P, Eklund A, Linden C. Pascal, ICare and Goldmann applanation tonometry--a comparative study. Acta Ophthalmol. 2008;86(6):614–621. - PubMed

Publication types