Reliability of goniometric measurements in children with cerebral palsy: a comparative analysis of universal goniometer and electronic inclinometer. A pilot study
- PMID: 21740600
- PMCID: PMC3160434
- DOI: 10.1186/1471-2474-12-155
Reliability of goniometric measurements in children with cerebral palsy: a comparative analysis of universal goniometer and electronic inclinometer. A pilot study
Abstract
Background: Even though technological progress has provided us with more and more sophisticated equipment for making goniometric measurements, the most commonly used clinical tools are still the universal goniometer and, to a lesser extent, the inclinometer. There is, however, no published study so far that uses an inclinometer for measurements in children with cerebral palsy (CP). The objective of this study was two-fold: to independently assess the intra and inter-examiner reliability for measuring the hip abduction range of motion in children with CP using two different instruments, the universal two-axis goniometer and electronic inclinometer. A pool of 5 examiners with different levels of experience as paediatric physiotherapists participated. The study did not compare both instruments because the measurement procedure and the hip position were different for each.
Methods: A prospective, observational study of goniometery was carried out with 14 lower extremities of 7 children with spastic CP. The inclinometer study was carried out with 8 lower extremities of 4 children with spastic CP. This study was divided into two independent parts: a study of the reliability of the hip abduction range of motion measured with a universal goniometer (hip at 0° flexion) and with an electronic inclinometer (hip at 90° flexion). The Intraclass Correlation Coefficient (ICC) was calculated to analyse intra and inter-examiner agreement for each instrument.
Results: For the goniometer, the intra-examiner reliability was excellent (>0.80), while the inter-examiner reliability was low (0.375 and 0.475). For the inclinometer, both the intra-examiner (0.850-0.975) and inter-examiner reliability were excellent (0.965 and 0.979).
Conclusions: The inter-examiner reliability for goniometric measurement of hip abduction in children with CP was low, in keeping with other results found in previous publications. The inclinometer has proved to be a highly reliable tool for measuring the hip abduction range of motion in children with CP, which opens up new possibilities in this field, despite having some measurement limitations.
Figures


Similar articles
-
Comparison of Functional Ankle Motion Measures in Modern Dancers.J Dance Med Sci. 2012 Sep;16(3):116-25. J Dance Med Sci. 2012. PMID: 26730940
-
Reliability of goniometric measurements in children with spastic cerebral palsy.Med Sci Monit. 2007 Jul;13(7):CR323-9. Med Sci Monit. 2007. PMID: 17599027
-
Reliability of inclinometer and goniometric measurements of hip extension flexibility using the modified Thomas test.Physiother Theory Pract. 2008 Mar-Apr;24(2):135-41. doi: 10.1080/09593980701378256. Physiother Theory Pract. 2008. PMID: 18432516
-
Intratester and intertester reliability and criterion validity of the parallelogram and universal goniometers for active knee flexion in healthy subjects.Physiother Res Int. 1997;2(3):150-66. doi: 10.1002/pri.97. Physiother Res Int. 1997. PMID: 9421820 Review.
-
The reliability and validity of goniometric elbow measurements in adults: A systematic review of the literature.Shoulder Elbow. 2018 Oct;10(4):274-284. doi: 10.1177/1758573218774326. Epub 2018 Jun 3. Shoulder Elbow. 2018. PMID: 30214494 Free PMC article. Review.
Cited by
-
Concurrent validity of digital inclinometer and universal goniometer in assessing passive hip mobility in healthy subjects.Int J Sports Phys Ther. 2013 Oct;8(5):680-8. Int J Sports Phys Ther. 2013. PMID: 24175147 Free PMC article.
-
Interrater reliability for unilateral and bilateral tests to measure the popliteal angle in children and youth with cerebral palsy.BMC Musculoskelet Disord. 2021 Mar 13;22(1):275. doi: 10.1186/s12891-021-04135-6. BMC Musculoskelet Disord. 2021. PMID: 33714264 Free PMC article.
-
Determining the Reliability of a New Method for Measuring Joint Range of Motion Through a Randomized Controlled Trial.Ann Rehabil Med. 2019 Dec;43(6):707-719. doi: 10.5535/arm.2019.43.6.707. Epub 2019 Dec 31. Ann Rehabil Med. 2019. PMID: 31918533 Free PMC article.
-
Validity, Reliability, and Efficiency of a Standard Goniometer, Medical Inclinometer, and Builder's Inclinometer.Int J Sports Phys Ther. 2023 Aug 1;18(4):989-996. doi: 10.26603/001c.83944. eCollection 2023. Int J Sports Phys Ther. 2023. PMID: 37547826 Free PMC article.
-
Application of Inertial Measurement Units and Machine Learning Classification in Cerebral Palsy: Randomized Controlled Trial.JMIR Rehabil Assist Technol. 2021 Oct 20;8(4):e29769. doi: 10.2196/29769. JMIR Rehabil Assist Technol. 2021. PMID: 34668870 Free PMC article.
References
-
- Hainsworth F, Harrison MJ, Sheldon TA, Roussounis SH. A preliminary evaluation of ankle orthoses in the management of children with cerebral palsy. Dev Med Child Neurol. 1997;39:243–247. - PubMed
-
- Koman LA, Brashear A, Rosenfeld S, Chambers H, Russman B, Rang M, Root L, Ferrari E, Garcia de Yebenes Prous J, Smith BP, Turkel C, Walcott JM, Molloy PT. Botulinum toxin type a neuromuscular blockade in the treatment of equinus foot deformity in cerebral palsy: a multicenter, open-label clinical trial. Pediatrics. 2001;108:1062–1071. doi: 10.1542/peds.108.5.1062. - DOI - PubMed
-
- Sala DA, Grant AD, Kummer FJ. Equinus deformity in cerebral palsy: recurrence after tendo Achillis lengthening. Dev Med Child Neurol. 1997;39:45–48. - PubMed
-
- Atar D, Zilberberg L, Votemberg M, Norsy M, Galil A. Effect of distal hamstrings release on cerebral palsy patients. Bulletin Hospital Joint Dis. 1993;53:34–36. - PubMed
Publication types