Superficial dose distribution in breast for tangential radiation treatment, Monte Carlo evaluation of Eclipse algorithms in case of phantom and patient geometries
- PMID: 21741719
- DOI: 10.1016/j.radonc.2011.06.021
Superficial dose distribution in breast for tangential radiation treatment, Monte Carlo evaluation of Eclipse algorithms in case of phantom and patient geometries
Abstract
Purpose: The aim of this study is to examine experimentally and by the Monte Carlo method the accuracy of the Eclipse Pencil Beam Convolution (PBC) and Analytical Anisotropic Algorithm (AAA) algorithms in the superficial region (0-2 cm) of the breast for tangential photon beams in a phantom case as well as in a number of patient geometries. The aim is also to identify differences in how the patient computer tomography data are handled by the treatment planning system and in the Monte Carlo simulations in order to reduce influences of these effects on the evaluation.
Materials and methods: Measurements by thermoluminescent dosimeters and gafchromic film are performed for six MV tangential irradiation of the cylindrical solid water phantom. Tangential treatment of seven patients is investigated considering open beams. Dose distributions are obtained by the Eclipse PBC and AAA algorithms. Monte Carlo calculations are carried out by BEAMnrc/DOSXYZnrc code package. Calculations are performed with a calculation grid of 1.25×1.25×5 mm(3) for PBC and 2×2×5 mm(3) for AAA and Monte Carlo, respectively. Dose comparison is performed in both dose and spatial domains by the normalized dose difference method.
Results: Experimental profiles from the surface toward the geometrical center of the cylindrical phantom are obtained at the beam entrance and exit as well as laterally. Full dose is received beyond 2 mm in the lateral superficial region and beyond 7 mm at the beam entrance. Good agreement between experimental, Monte Carlo and AAA data is obtained, whereas PBC is seen to underestimate the entrance dose the first 3-4 mm and the lateral dose by more than 5% up to 8 mm depth. In the patient cases considered, AAA and Monte Carlo show agreement within 3% dose and 4 mm spatial tolerance. PBC systematically underestimates the dose at the breast apex. The dimensions of region out of tolerance vary with the local breast shape. Different interpretations of patient boundaries in Monte Carlo and the Eclipse are found to influence the evaluation. Computer tomography marker wire may introduce local disturbance effects on the comparison as well. These factors are not related to the accuracy of the calculation algorithms and their effect is taken into account in the evaluation.
Conclusions: The accuracy of AAA in the case of the solid water phantom is comparable with that of the Monte Carlo method. The AAA-Monte Carlo differences in the patient cases considered are within 3%, 4 mm tolerance. The PBC algorithm does not give equivalent results. In the phantom case, PBC underestimates the lateral dose by more than 5% up to 8 mm depth. The PBC-Monte Carlo differences in the patient cases are outside the tolerance at the breast apex. The dimension of region varies with the breast shape being typically 8-10 mm long and 6-8 mm deep.
Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.
Similar articles
-
AAA and PBC calculation accuracy in the surface build-up region in tangential beam treatments. Phantom and breast case study with the Monte Carlo code PENELOPE.Radiother Oncol. 2009 Oct;93(1):94-101. doi: 10.1016/j.radonc.2009.05.010. Epub 2009 Jun 21. Radiother Oncol. 2009. PMID: 19541380
-
Advanced kernel methods vs. Monte Carlo-based dose calculation for high energy photon beams.Radiother Oncol. 2009 Dec;93(3):645-53. doi: 10.1016/j.radonc.2009.10.013. Epub 2009 Nov 18. Radiother Oncol. 2009. PMID: 19926153
-
The impact of photon dose calculation algorithms on expected dose distributions in lungs under different respiratory phases.Phys Med Biol. 2008 May 7;53(9):2375-90. doi: 10.1088/0031-9155/53/9/011. Epub 2008 Apr 17. Phys Med Biol. 2008. PMID: 18421117
-
High-density dental implants and radiotherapy planning: evaluation of effects on dose distribution using pencil beam convolution algorithm and Monte Carlo method.J Appl Clin Med Phys. 2015 Sep 8;16(5):46–52. doi: 10.1120/jacmp.v16i5.5612. J Appl Clin Med Phys. 2015. PMID: 26699323 Free PMC article. Review.
-
Optimization of the Eclipse TPS beam configuration parameters for small field dosimetry using Monte Carlo simulations and experimental measurements.Phys Med. 2023 Oct;114:103141. doi: 10.1016/j.ejmp.2023.103141. Epub 2023 Oct 30. Phys Med. 2023. PMID: 37820506 Review.
Cited by
-
Dosimetric comparison of preoperative single-fraction partial breast radiotherapy techniques: 3D CRT, noncoplanar IMRT, coplanar IMRT, and VMAT.J Appl Clin Med Phys. 2015 Jan 8;16(1):5126. doi: 10.1120/jacmp.v16i1.5126. J Appl Clin Med Phys. 2015. PMID: 25679170 Free PMC article.
-
Optimization of skin dose using in-vivo MOSFET dose measurements in bolus/non-bolus fraction ratio: A VMAT and a 3DCRT study.J Appl Clin Med Phys. 2019 Feb;20(2):63-70. doi: 10.1002/acm2.12525. Epub 2019 Jan 9. J Appl Clin Med Phys. 2019. PMID: 30628154 Free PMC article.
-
Skin recurrence in the radiation treatment of breast cancer.Adv Radiat Oncol. 2018 May 7;3(3):458-462. doi: 10.1016/j.adro.2018.04.014. eCollection 2018 Jul-Sep. Adv Radiat Oncol. 2018. PMID: 30202813 Free PMC article. No abstract available.
-
Comparing the monoisocentric and dual isocentric techniques in chest wall radiotherapy of mastectomy patients.J Appl Clin Med Phys. 2015 Jan 8;16(1):5069. doi: 10.1120/jacmp.v16i1.5069. J Appl Clin Med Phys. 2015. PMID: 25679164 Free PMC article.
-
Clinical implementation of AXB from AAA for breast: Plan quality and subvolume analysis.J Appl Clin Med Phys. 2018 May;19(3):243-250. doi: 10.1002/acm2.12329. Epub 2018 Apr 25. J Appl Clin Med Phys. 2018. PMID: 29696752 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous