Image quality and radiation dose comparison of prospectively triggered low-dose CCTA: 128-slice dual-source high-pitch spiral versus 64-slice single-source sequential acquisition
- PMID: 21744246
- DOI: 10.1007/s10554-011-9921-3
Image quality and radiation dose comparison of prospectively triggered low-dose CCTA: 128-slice dual-source high-pitch spiral versus 64-slice single-source sequential acquisition
Abstract
Currently 64-multislice computed tomography (MSCT) scanners are the most widely used devices allowing low radiation dose coronary CT angiography (CCTA) with prospective ECG triggering. Latest 128-slice dual-source CT (DSCT) scanners offer prospective high-pitch spiral acquisition covering the heart during one single beat. We compared radiation dose and image quality from prospective 64-MSCT versus high-pitch spiral 128-slice DSCT scanning, as such data is lacking. CCTA of 50 consecutive patients undergoing 128-DSCT (2 × 64 × 0.6 mm collimation, 0.28 s rotation time, 3.4 pitch, 100-120 kV tube voltage and 320 mAs tube current-time product) were compared to CCTA of 50 heart rate (HR) and BMI matched patients undergoing 64-MSCT (64 × 0.625 mm collimation, 0.35 s rotation time, 100-120 kV tube voltage and 400-650 mA tube current). Image quality was rated on a 4-point scale by two independent cardiac imaging physicians (1 = excellent to 4 = non-diagnostic). Of 710 coronary segments assessed on 128-DSCT, 216 (30.4%) achieved an image quality score 1 excellent, 400 (56.3%) score 2, 76 (10.7%) score 3 and 18 (2.6%) score 4 (non-diagnostic). Of 737 coronary segments evaluated on 64-MSCT 271 (36.8%) had an image quality score of 1, 327 (44.4%) 2, 110 (14.9%) score 3, and 29 (3.9%) segments score 4. Average image quality score for both scanners was similar (P = 0.641). The mean heart rate during scanning was 58.7 ± 5.6 bpm on 128-DSCT and 59.0 ± 5.6 bpm on 64-MSCT, respectively. Mean effective radiation dose was 1.0 ± 0.2 mSv for 128-DSCT and 1.7 ± 0.6 mSv for 64-MSCT (P < 0.001). 128-DSCT with high-pitch spiral mode allows CCTA acquisition with reduced radiation dose at maintained image quality compared to 64-MSCT.
Similar articles
-
Prospectively ECG-triggered high-pitch coronary angiography with third-generation dual-source CT at 70 kVp tube voltage: feasibility, image quality, radiation dose, and effect of iterative reconstruction.J Cardiovasc Comput Tomogr. 2014 Nov-Dec;8(6):418-25. doi: 10.1016/j.jcct.2014.09.003. Epub 2014 Sep 16. J Cardiovasc Comput Tomogr. 2014. PMID: 25439789
-
Prospectively versus retrospectively ECG-gated 256-slice coronary CT angiography: image quality and radiation dose over expanded heart rates.Int J Cardiovasc Imaging. 2012 Jan;28(1):153-62. doi: 10.1007/s10554-010-9760-7. Epub 2010 Dec 14. Int J Cardiovasc Imaging. 2012. PMID: 21153709
-
Performance of turbo high-pitch dual-source CT for coronary CT angiography: first ex vivo and patient experience.Eur Radiol. 2014 Aug;24(8):1889-95. doi: 10.1007/s00330-014-3209-7. Epub 2014 May 17. Eur Radiol. 2014. PMID: 24838737
-
Coronary computed tomography angiography with prospective electrocardiography triggering: a systematic review of image quality and radiation dose.Singapore Med J. 2013 Jan;54(1):15-23. doi: 10.11622/smedj.2013005. Singapore Med J. 2013. PMID: 23338911
-
Acquisition and Reconstruction Techniques for Coronary CT Angiography: Current Status and Trends over the Past Decade.Radiographics. 2025 Jul;45(7):e240083. doi: 10.1148/rg.240083. Radiographics. 2025. PMID: 40504731 Review.
Cited by
-
Congenital coronary anomalies detected by coronary computed tomography compared to invasive coronary angiography.BMC Cardiovasc Disord. 2014 Jul 8;14:81. doi: 10.1186/1471-2261-14-81. BMC Cardiovasc Disord. 2014. PMID: 25004927 Free PMC article.
-
First in vivo head-to-head comparison of high-definition versus standard-definition stent imaging with 64-slice computed tomography.Int J Cardiovasc Imaging. 2013 Aug;29(6):1409-16. doi: 10.1007/s10554-013-0225-7. Epub 2013 May 1. Int J Cardiovasc Imaging. 2013. PMID: 23636300
-
CT coronary angiography: impact of adapted statistical iterative reconstruction (ASIR) on coronary stenosis and plaque composition analysis.Int J Cardiovasc Imaging. 2013 Mar;29(3):719-24. doi: 10.1007/s10554-012-0134-1. Epub 2012 Oct 2. Int J Cardiovasc Imaging. 2013. PMID: 23053859
-
Diagnostic accuracy of coronary CT angiography combined with dual-energy myocardial perfusion imaging for detection of myocardial infarction.Exp Ther Med. 2017 Jul;14(1):207-213. doi: 10.3892/etm.2017.4485. Epub 2017 May 22. Exp Ther Med. 2017. PMID: 28672916 Free PMC article.
-
Computed Tomography Angiography with a 192-slice Dual-source Computed Tomography System: Improvements in Image Quality and Radiation Dose.J Clin Imaging Sci. 2016 Oct 21;6:44. doi: 10.4103/2156-7514.192840. eCollection 2016. J Clin Imaging Sci. 2016. PMID: 27833784 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical