. 2011 Nov;38(11):2025-30.
doi: 10.1007/s00259-011-1877-y. Epub 2011 Jul 15.

Nuclear myocardial perfusion imaging with a novel cadmium-zinc-telluride detector SPECT/CT device: first validation versus invasive coronary angiography

Affiliations

Nuclear myocardial perfusion imaging with a novel cadmium-zinc-telluride detector SPECT/CT device: first validation versus invasive coronary angiography

Michael Fiechter et al. Eur J Nucl Med Mol Imaging. 2011 Nov.

Abstract

Purpose: We evaluated the diagnostic accuracy of attenuation corrected nuclear myocardial perfusion imaging (MPI) with a novel hybrid single photon emission computed tomography (SPECT)/CT device consisting of an ultrafast dedicated cardiac gamma camera with cadmium-zinc-telluride (CZT) solid-state semiconductor detectors integrated onto a multislice CT scanner to detect coronary artery disease (CAD). Invasive coronary angiography served as the standard of reference.

Methods: The study population included 66 patients (79% men; mean age 63 ± 11 years) who underwent 1-day (99m)Tc-tetrofosmin pharmacological stress/rest examination and angiography within 3 months. Sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) as well as accuracy of the CT X-ray based attenuation corrected CZT MPI for detection of CAD (≥ 50% luminal narrowing) was calculated on a per-patient basis.

Results: The prevalence of angiographic CAD in the study population was 82%. Sensitivity, specificity, PPV, NPV and accuracy were 87, 67, 92, 53 and 83%, respectively.

Conclusion: In this first report on CZT SPECT/CT MPI comparison versus angiography we confirm a high accuracy for detection of angiographically documented CAD.

PubMed Disclaimer

References

    1. J Nucl Cardiol. 2010 Jan-Feb;17(1):4-7 - PubMed
    1. Eur J Nucl Med Mol Imaging. 2010 Nov;37(11):2086-92 - PubMed
    1. J Am Coll Cardiol. 1993 Nov 1;22(5):1455-64 - PubMed
    1. N Engl J Med. 1983 Sep 1;309(9):518-22 - PubMed
    1. Eur J Nucl Med Mol Imaging. 2010 Oct;37(10):1903-8 - PubMed

Publication types

MeSH terms