Pulmonary arterial thermodilution, femoral arterial thermodilution and bioreactance cardiac output monitoring in a pediatric hemorrhagic hypovolemic shock model
- PMID: 21763249
- DOI: 10.1016/j.resuscitation.2011.06.039
Pulmonary arterial thermodilution, femoral arterial thermodilution and bioreactance cardiac output monitoring in a pediatric hemorrhagic hypovolemic shock model
Abstract
Aim: Bioreactance is a new non-invasive method for cardiac output measurement (NICOM). There are no studies that have analysed the utility of this technique in a pediatric animal model of hemorrhagic shock.
Methods: A prospective study was performed using 9 immature Maryland pigs weighing 9 to 12 kg was performed. A Swan-Ganz catheter, a PiCCO catheter and 4 dual surface electrodes were placed at the four corners of the anterior thoracic body surface. Shock was induced by withdrawing a blood volume of 30 mL/kg, and then after, 30 mL/kg of Normal saline was administered. Seven simultaneous measurements of cardiac index (CI) were made by pulmonary artery thermodilution (PATD), Femoral artery thermodilution (FATD), and NICOM before, during, and after hypovolaemia and during and after volume expansion.
Results: The mean difference (bias) of differences (limits of agreement) between PATD and FATD was 0.84 (-1.87-3.51)L/min/1.77 m(2), between PATD and NICOM was 1.95 (-1.79-5.69)L/min/1.77 m(2), and between FATD and NICOM was 1.06 (-1.40-3.52)L/min/1.77 m(2). A moderate correlation was found between PATD and FATD (r=0.43; P=0.01), but no correlation was found between bioreactance and either PATD or FATD. Hypovolemia and volume expansion produced important significant differences in CI as measured by PATD and FATD, while the changes with bioreactance were small and non significant.
Conclusions: PATD and FATD measurements showed similar responses to hypovolemic shock and volume expansion. Bioreactance persistently underestimates the CI and is not significantly altered by either inducing hemorrhagic shock, or later, through volume expansion. Bioreactance is not a suitable method for monitoring the CI in pediatric hemorrhagic shock.
Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.
Similar articles
-
Pulse contour cardiac output analysis in a piglet model of severe hemorrhagic shock.Crit Care Med. 2008 Apr;36(4):1189-95. doi: 10.1097/CCM.0B013E31816592A3. Crit Care Med. 2008. PMID: 18401255
-
Comparison between cardiac output measured by the pulmonary arterial thermodilution technique and that measured by the femoral arterial thermodilution technique in a pediatric animal model.Pediatr Cardiol. 2004 Mar-Apr;25(2):119-23. doi: 10.1007/s00246-003-0450-2. Epub 2003 Dec 23. Pediatr Cardiol. 2004. PMID: 14681741
-
Agreement of Bioreactance Cardiac Output Monitoring With Thermodilution During Hemorrhagic Shock and Resuscitation in Adult Swine.Crit Care Med. 2017 Feb;45(2):e195-e201. doi: 10.1097/CCM.0000000000002071. Crit Care Med. 2017. PMID: 27749345
-
Agreement between continuous and intermittent pulmonary artery thermodilution for cardiac output measurement in perioperative and intensive care medicine: a systematic review and meta-analysis.Crit Care. 2021 Mar 29;25(1):125. doi: 10.1186/s13054-021-03523-7. Crit Care. 2021. PMID: 33781307 Free PMC article.
-
Accuracy and precision of non-invasive cardiac output monitoring by electrical cardiometry: a systematic review and meta-analysis.J Clin Monit Comput. 2020 Jun;34(3):433-460. doi: 10.1007/s10877-019-00330-y. Epub 2019 Jun 7. J Clin Monit Comput. 2020. PMID: 31175501 Free PMC article.
Cited by
-
Non-invasive Cardiac Output Monitoring and Assessment of Fluid Responsiveness in Children With Shock in the Emergency Department.Front Pediatr. 2022 Apr 7;10:857106. doi: 10.3389/fped.2022.857106. eCollection 2022. Front Pediatr. 2022. PMID: 35463892 Free PMC article.
-
Cardiac index assessment using bioreactance in patients undergoing cytoreductive surgery in ovarian carcinoma.J Clin Monit Comput. 2013 Dec;27(6):621-7. doi: 10.1007/s10877-013-9478-x. Epub 2013 May 21. J Clin Monit Comput. 2013. PMID: 23689837 Clinical Trial.
-
Comparison between pressure-recording analytical method (PRAM) and femoral arterial thermodilution method (FATD) cardiac output monitoring in an infant animal model of cardiac arrest.Intensive Care Med Exp. 2016 Dec;4(1):13. doi: 10.1186/s40635-016-0087-0. Epub 2016 Jun 3. Intensive Care Med Exp. 2016. PMID: 27256288 Free PMC article.
-
Measurement of cardiac output in children by pressure-recording analytical method.Pediatr Cardiol. 2015 Feb;36(2):358-64. doi: 10.1007/s00246-014-1014-3. Epub 2014 Sep 2. Pediatr Cardiol. 2015. PMID: 25179459
-
Comparison of monitoring performance of bioreactance versus esophageal Doppler in pediatric patients.Indian J Crit Care Med. 2015 Jan;19(1):3-8. doi: 10.4103/0972-5229.148630. Indian J Crit Care Med. 2015. PMID: 25624643 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources