A comparative biomechanical analysis of fixation devices for unstable femoral neck fractures: the Intertan versus cannulated screws or a dynamic hip screw
- PMID: 21768904
- DOI: 10.1097/TA.0b013e31820e86e6
A comparative biomechanical analysis of fixation devices for unstable femoral neck fractures: the Intertan versus cannulated screws or a dynamic hip screw
Abstract
Background: First introduced in 2005, the "Intertan" (IT), an intramedullary nail with two cephalocervical screws, has become an increasingly popular option for treating intertrochanteric fractures. The purpose of this study was to identify the utility of this device for stabilization of unstable femoral neck fractures compared with cannulated screws (CS) and a dynamic hip screw (DHS).
Methods: Twenty-four human cadaveric femurs were harvested and assigned to three groups that were matched with regard to bone mineral density (BMD). Standardized Pauwels-Type-III fractures were osteomized with a custom-made saw guide and fixated by an "IT," three CS, or a DHS. All constructs were biomechanically tested in a servohydraulic testing machine with a physiologic mechanical axis loading of the femoral head (700 N), cyclical compression to 1,400 N (10,000 cycles; 2 Hz), and loading to failure. All specimens were compared with respect to the number of survived cycles, mechanical strength, head displacement, load to failure, and failure mechanism.
Results: Regardless of the fixation, the mechanical strength of the stabilized femurs was significantly decreased to 71% compared with the intact femurs (100%). During cyclical testing 46% of the constructs (6 CS, 4 DHS, and 1 IT) failed. There was no difference between the mechanical strength of all survived constructs regarding the BMD, but the BMD of the failed specimens was significantly reduced compared with the surviving femurs (0.71 g/cm² ± 0.18 g/cm² vs. 1.07 g/cm² ± 0.33 g/cm²; p < 0.05). IT femurs survived significantly longer than CS specimens (IT, 9,063 cycles ± 2,480 cycles vs. CS, 3,325 cycles ± 3,885 cycles vs. DHS, 5,716 cycles ± 4,448 cycles; p < 0.01), endured higher failure loads (IT, 4,929 N ± 1,105 N vs. CS, 3,421 N ± 20 N vs. DHS, 3,505 N ± 905 N; p < 0.05), and presented a less inferior head displacement (IT, 8.5 mm ± 1.6 mm vs. CS, 16.4 mm ± 6.7 mm vs. DHS, 14.5 mm ± 6.4 mm; p < 0.05).
Conclusion: Our results suggest that (1) none of the tested devices restore a comparable mechanical strength in the fractured specimens compared with the intact femurs, and (2) the "IT" possesses some biomechanical benefits for internal fixation of unstable femoral neck fractures compared with DHS and CS. Because the IT constructs failed with an inferior femoral neck fracture, complicating the mandatory anchorage of a prosthetic stem in a revision operation, more biomechanical experiments using the IT in the presence of a posterior comminution defect are required, along with clinical outcome studies.
Similar articles
-
Internal fixation of femoral neck fractures with posterior comminution: a biomechanical comparison of DHS® and Intertan nail®.Int Orthop. 2011 Nov;35(11):1695-701. doi: 10.1007/s00264-010-1199-x. Epub 2011 Feb 1. Int Orthop. 2011. PMID: 21286918 Free PMC article.
-
Vertically oriented femoral neck fractures: mechanical analysis of four fixation techniques.J Orthop Trauma. 2007 Sep;21(8):544-8. doi: 10.1097/BOT.0b013e31814b822e. J Orthop Trauma. 2007. PMID: 17805021
-
Is a helical shaped implant a superior alternative to the Dynamic Hip Screw for unstable femoral neck fractures? A biomechanical investigation.Clin Biomech (Bristol). 2009 Jan;24(1):59-64. doi: 10.1016/j.clinbiomech.2008.07.004. Epub 2008 Oct 31. Clin Biomech (Bristol). 2009. PMID: 18977058
-
Biomechanical Evaluation of Internal Fixation of Pauwels Type III Femoral Neck Fractures: A Systematic Review of Various Fixation Methods.Clin Orthop Surg. 2019 Mar;11(1):1-14. doi: 10.4055/cios.2019.11.1.1. Epub 2019 Feb 18. Clin Orthop Surg. 2019. PMID: 30838102 Free PMC article.
-
Multiple cannulated screws vs. dynamic hip screws for femoral neck fractures : A meta-analysis.Orthopade. 2017 Nov;46(11):954-962. doi: 10.1007/s00132-017-3473-8. Orthopade. 2017. PMID: 29022057 Review. English.
Cited by
-
Efficacy of the femoral neck system in femoral neck fracture treatment in adults: A systematic review and meta-analysis.World J Clin Cases. 2022 Nov 6;10(31):11454-11465. doi: 10.12998/wjcc.v10.i31.11454. World J Clin Cases. 2022. PMID: 36387785 Free PMC article.
-
Finite element analysis of different configurations of fully threaded cannulated screw in the treatment of unstable femoral neck fractures.J Orthop Surg Res. 2018 Oct 29;13(1):272. doi: 10.1186/s13018-018-0970-3. J Orthop Surg Res. 2018. PMID: 30373617 Free PMC article.
-
Comparison of three different internal fixation implants in treatment of femoral neck fracture-a finite element analysis.J Orthop Surg Res. 2019 Mar 12;14(1):76. doi: 10.1186/s13018-019-1097-x. J Orthop Surg Res. 2019. PMID: 30871584 Free PMC article.
-
A case of total hip arthroplasty for nonunion after femoral trochanteric fracture surgery with complete breakage of only the compression screw of the InterTAN nail.Trauma Case Rep. 2023 Sep 23;48:100939. doi: 10.1016/j.tcr.2023.100939. eCollection 2023 Dec. Trauma Case Rep. 2023. PMID: 37810539 Free PMC article.
-
InterTan nail versus Gamma3 nail for intramedullary nailing of unstable trochanteric fractures.Diagn Pathol. 2014 Oct 1;9:191. doi: 10.1186/s13000-014-0191-y. Diagn Pathol. 2014. PMID: 25269555 Free PMC article.