Mid-term results of total hip arthroplasty with the CementLess Spotorno (CLS) system
- PMID: 21776579
- DOI: 10.2106/JBJS.I.01792
Mid-term results of total hip arthroplasty with the CementLess Spotorno (CLS) system
Abstract
Background: Except for those reported by the designers, there are no published mid-term results of the use of the CementLess Spotorno (CLS) Total Hip Arthroplasty system. We present the results of (1) a ten to seventeen-year follow-up prospective cohort study of this system, and (2) retrospective analyses of factors influencing clinical and radiographic outcomes.
Methods: We studied a series of 102 consecutive CLS arthroplasties with a minimal duration of follow-up of ten years. Indications for the procedures were osteoarthritis (n = 90), rheumatoid arthritis (n = 8), and femoral head osteonecrosis (n = 4). The Merle d'Aubigné-Postel score, polyethylene wear, and radiographic status were recorded at regular intervals. Survival analyses, repeated-measures analysis of variance, and a nested case-control study (with the cases having early revision due to aseptic cup loosening within ten years after the index procedure and the controls having no early cup revision) were used for evaluation.
Results: There were fourteen revisions, including nine due to aseptic cup loosening. The ten-year Kaplan-Meier survival rate was 92.2% (95% confidence interval [CI] = 86.9 to 97.5) with revision for any reason as the end point. The fifteen-year survival rate was 78.4% (95% CI = 63.9 to 92.9) with revision for any reason as the end point, 81.6% (95% CI = 66.7 to 96.5) with revision due to aseptic cup loosening as the end point, and 99.0% (95% CI = 97.0 to 100.0) with revision due to aseptic stem loosening as the end point. The average amount of polyethylene wear at the time of final follow-up was 1.92 mm (range, 0.6 to 4.3 mm). The wear rate in the cases was significantly higher than that in the controls (0.31 vs. 0.16 mm/yr, p < 0.001). Factors with a significant effect on polyethylene wear were age at surgery (a 0.3-mm increase per every ten years younger, p = 0.001) and a larger head component (an effect of 0.53 mm for the 32 vs. the 28-mm component; p < 0.0001). Male sex had an effect of -0.66 point (p = 0.07) on the final Merle d'Aubigné-Postel score.
Conclusions: The results of this CLS system, particularly with regard to the femoral stem, are comparable with those with other reliable cementless systems. Nevertheless, the prevalence of aseptic acetabular cup loosening in the second decade after the operation demonstrates a potentially substantial problem with regard to long-term survival. A high polyethylene wear rate, male sex, a younger age at the time of surgery, and a 32-mm head component size are related to inferior clinical outcomes and a higher risk of implant revision.
Similar articles
-
Long-term results of total hip arthroplasty with the CementLess Spotorno (CLS) system.Hip Int. 2017 Sep 19;27(5):465-471. doi: 10.5301/hipint.5000492. Epub 2017 May 27. Hip Int. 2017. PMID: 28574121
-
Poor results of the cementless total hip arthroplasty with a nonporous coated acetabular component: AcSys Shearer Cup.J Orthop Sci. 2007 Sep;12(5):466-70. doi: 10.1007/s00776-007-1161-y. Epub 2007 Sep 28. J Orthop Sci. 2007. PMID: 17909932
-
Total hip arthroplasty survival in femoral head avascular necrosis versus primary hip osteoarthritis: Case-control study with a mean 10-year follow-up after anatomical cementless metal-on-metal 28-mm replacement.Orthop Traumatol Surg Res. 2016 Dec;102(8):1029-1034. doi: 10.1016/j.otsr.2016.08.021. Epub 2016 Oct 27. Orthop Traumatol Surg Res. 2016. PMID: 28341264
-
Performance of the CLS Spotorno uncemented stem in the third decade after implantation.Bone Joint J. 2014 Apr;96-B(4):455-61. doi: 10.1302/0301-620X.96B4.32607. Bone Joint J. 2014. PMID: 24692610 Review.
-
Over ten-year follow-up results of a prospective and consecutive series of primary total hip arthroplasty with an original cementless total hip prosthesis.Int Orthop. 2024 Apr;48(4):945-954. doi: 10.1007/s00264-023-06071-7. Epub 2023 Dec 28. Int Orthop. 2024. PMID: 38153431 Free PMC article. Review.
Cited by
-
Long-term (20- to 25-year) results of an uncemented tapered titanium femoral component and factors affecting survivorship.Clin Orthop Relat Res. 2013 Oct;471(10):3262-9. doi: 10.1007/s11999-013-3033-4. Epub 2013 May 14. Clin Orthop Relat Res. 2013. PMID: 23670671 Free PMC article.
-
Outcome of the Cementless Zweymüller BICON-PLUS Cup and SL-PLUS Stem in the Very Elderly Individuals.Geriatr Orthop Surg Rehabil. 2016 Jun;7(2):74-80. doi: 10.1177/2151458516638902. Epub 2016 Mar 24. Geriatr Orthop Surg Rehabil. 2016. PMID: 27239380 Free PMC article.
-
Twenty-year results of a neck-preserving short-stem prosthesis in primary total hip arthroplasty.Arch Orthop Trauma Surg. 2023 Jun;143(6):3481-3486. doi: 10.1007/s00402-022-04556-5. Epub 2022 Jul 29. Arch Orthop Trauma Surg. 2023. PMID: 35906493
-
Second-generation uncemented total hip arthroplasty: a minimum 20-year follow-up.Bone Jt Open. 2021 Jan 13;2(1):33-39. doi: 10.1302/2633-1462.21.BJO-2020-0157.R1. eCollection 2021 Jan. Bone Jt Open. 2021. PMID: 33537674 Free PMC article.
-
Mid-term results of an uncemented tapered femoral stem and various factors affecting survivorship.J Clin Orthop Trauma. 2019 Mar-Apr;10(2):368-373. doi: 10.1016/j.jcot.2018.02.001. Epub 2018 Feb 5. J Clin Orthop Trauma. 2019. PMID: 30828210 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical