Abstract
Background: Two-piece implants present gaps and cavities between the implant and the abutment, and these hollow spaces can act as a trap for bacteria. The aim of the present in vitro study was to evaluate the migration of two different microbial species from the inside to the outside of the implant-abutment assembly in three different connection types.
Methods: A total of 30 implants (10 implants per group) were used. The implants presented a screwed trilobed connection (group 1), a cemented connection (group 2), and an internal conical connection (group 3). The inner parts of five implants, per group, were inoculated with Pseudomonas aeruginosa suspension and the remaining five implants, per group, with Aggregatibacter actinomycetemcomitans. The penetration of bacteria into the surrounding solution was determined by the observation of turbidity of the broth.
Results: In group 1, bacterial contamination was found in six of 10 implants. In group 2, no contaminated samples were found. In group 3, bacterial contamination was found in one implant of 10. Statistically significant differences were detected between group 1 versus group 3 (P <0.05) and between group 1 versus group 2 (P <0.01), whereas no significant differences were found when comparing group 2 versus group 3 (P >0.05).
Conclusion: The present study confirms previous results about the hermeticity of the cement-retained implant-abutment assembly, the very low permeability to bacteria of the conical implant-abutment connection, and the high prevalence of bacterial penetration of screw-retained implant-abutment assemblies.
Similar articles
-
An in vitro investigation concerning the bacterial leakage at implants with internal hexagon and Morse taper implant-abutment connections.Implant Dent. 2012 Aug;21(4):335-9. doi: 10.1097/ID.0b013e31825cd472. Implant Dent. 2012. PMID: 22814560
-
Implants with internal hexagon and conical implant-abutment connections: an in vitro study of the bacterial contamination.J Oral Implantol. 2014 Feb;40(1):30-6. doi: 10.1563/AAID-JOI-D-11-00121. Epub 2011 Dec 30. J Oral Implantol. 2014. PMID: 22208157
-
Fluids and microbial penetration in the internal part of cement-retained versus screw-retained implant-abutment connections.J Periodontol. 2001 Sep;72(9):1146-50. doi: 10.1902/jop.2000.72.9.1146. J Periodontol. 2001. PMID: 11577944
-
[Microbial contamination of the implant-abutment connections: Review of the literature].Rev Stomatol Chir Maxillofac Chir Orale. 2016 Feb;117(1):20-5. doi: 10.1016/j.revsto.2015.11.004. Epub 2015 Dec 30. Rev Stomatol Chir Maxillofac Chir Orale. 2016. PMID: 26746601 Review. French.
-
Influence of methodologic aspects on the results of implant-abutment interface microleakage tests: a critical review of in vitro studies.Int J Oral Maxillofac Implants. 2012 Jul-Aug;27(4):793-800. Int J Oral Maxillofac Implants. 2012. PMID: 22848880 Review.
Cited by
-
Bacterial Microleakage at the Implant-Abutment Interface: An In Vitro Study.Bioengineering (Basel). 2022 Jun 26;9(7):277. doi: 10.3390/bioengineering9070277. Bioengineering (Basel). 2022. PMID: 35877328 Free PMC article.
-
3D X-ray microscopic analysys on a prosthetically loaded implant with platform-switching and conical connection: a case report.Oral Implantol (Rome). 2017 Nov 30;10(3):241-246. doi: 10.11138/orl/2017.10.3.241. eCollection 2017 Jul-Sep. Oral Implantol (Rome). 2017. PMID: 29285326 Free PMC article.
-
Evaluating the microgap and sealing capability in four implant systems with different interlockings under different tightening torques: an in-vitro study.J Adv Prosthodont. 2024 Dec;16(6):336-347. doi: 10.4047/jap.2024.16.6.336. Epub 2024 Dec 19. J Adv Prosthodont. 2024. PMID: 39803384 Free PMC article.
-
Determination of fluid leakages in the different screw-retained implant-abutment connections in a mechanical artificial mouth.J Mater Sci Mater Med. 2015 Jul;26(7):211. doi: 10.1007/s10856-015-5544-x. Epub 2015 Jul 15. J Mater Sci Mater Med. 2015. PMID: 26174348
-
Vertical Discrepancy in Height of Morse Cone Abutments Submitted to Different Torque Forces.Materials (Basel). 2021 Aug 30;14(17):4950. doi: 10.3390/ma14174950. Materials (Basel). 2021. PMID: 34501040 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources