Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2011 Oct;24(10):1079-85.
doi: 10.1016/j.echo.2011.06.011. Epub 2011 Jul 30.

Comparative accuracy of two- and three-dimensional transthoracic and transesophageal echocardiography in identifying mitral valve pathology in patients undergoing mitral valve repair: initial observations

Affiliations
Comparative Study

Comparative accuracy of two- and three-dimensional transthoracic and transesophageal echocardiography in identifying mitral valve pathology in patients undergoing mitral valve repair: initial observations

Sagit Ben Zekry et al. J Am Soc Echocardiogr. 2011 Oct.

Abstract

Background: Identification of mitral regurgitation (MR) mechanism and pathology are crucial for surgical repair. The aim of the present investigation was to evaluate the comparative accuracy of real-time three-dimensional (3D) transesophageal echocardiography (TEE) and transthoracic echocardiography (TTE) with two-dimensional (2D) TEE and TTE in diagnosing the mechanism of MR compared with the surgical standard.

Methods: Forty patients referred for surgical mitral valve repair were studied; 2D and 3D echocardiography with both TTE and TEE were performed preoperatively. Two independent observers reviewed the studies for MR pathology, functional or organic. In organic disease, the presence and localization of leaflet prolapse and/or flail were noted. Surgical findings served as the gold standard.

Results: There was 100% agreement in identifying functional versus organic MR among all four modalities. Overall, 2D TTE, 2D TEE, and 3D TEE performed similarly in identifying a prolapse or a flail leaflet; 3D TEE had the best agreement in identifying anterior leaflet prolapse, and it also showed an advantage for segmental analysis. Three-dimensional TTE was less sensitive and less accurate in identifying flail segments.

Conclusion: All modalities were equally reliable in identifying functional MR. Both 2D TEE and 3D TEE were comparable in diagnosing MR mechanism, while 3D TEE had the advantage of better localizing the disease. With current technology, 3D TTE was the least reliable in identifying valve pathology.

PubMed Disclaimer

Comment in

Similar articles

Cited by

Publication types

MeSH terms