Report of AAPM TG 135: quality assurance for robotic radiosurgery
- PMID: 21815366
- DOI: 10.1118/1.3579139
Report of AAPM TG 135: quality assurance for robotic radiosurgery
Erratum in
- Med Phys. 2011 Sep;38(9):5264
-
Erratum: "Report of AAPM TG 135: Quality assurance for robotic radiosurgery".Med Phys. 2011 Sep;38(9):5264. doi: 10.1118/1.3626480. Med Phys. 2011. PMID: 28524974 No abstract available.
Abstract
The task group (TG) for quality assurance for robotic radiosurgery was formed by the American Association of Physicists in Medicine's Science Council under the direction of the Radiation Therapy Committee and the Quality Assurance (QA) Subcommittee. The task group (TG-135) had three main charges: (1) To make recommendations on a code of practice for Robotic Radiosurgery QA; (2) To make recommendations on quality assurance and dosimetric verification techniques, especially in regard to real-time respiratory motion tracking software; (3) To make recommendations on issues which require further research and development. This report provides a general functional overview of the only clinically implemented robotic radiosurgery device, the CyberKnife. This report includes sections on device components and their individual component QA recommendations, followed by a section on the QA requirements for integrated systems. Examples of checklists for daily, monthly, annual, and upgrade QA are given as guidance for medical physicists. Areas in which QA procedures are still under development are discussed.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous