Pulse oximetry screening for congenital heart defects in newborn infants (PulseOx): a test accuracy study
- PMID: 21820732
- DOI: 10.1016/S0140-6736(11)60753-8
Pulse oximetry screening for congenital heart defects in newborn infants (PulseOx): a test accuracy study
Abstract
Background: Screening for congenital heart defects relies on antenatal ultrasonography and postnatal clinical examination; however, life-threatening defects often are not detected. We prospectively assessed the accuracy of pulse oximetry as a screening test for congenital heart defects.
Methods: In six maternity units in the UK, asymptomatic newborn babies (gestation >34 weeks) were screened with pulse oximetry before discharge. Infants who did not achieve predetermined oxygen saturation thresholds underwent echocardiography. All other infants were followed up to 12 months of age by use of regional and national registries and clinical follow-up. The main outcome was the sensitivity and specificity of pulse oximetry for detection of critical congenital heart defects (causing death or requiring invasive intervention before 28 days) or major congenital heart disease (causing death or requiring invasive intervention within 12 months of age).
Findings: 20,055 newborn babies were screened and 53 had major congenital heart disease (24 critical), a prevalence of 2·6 per 1000 livebirths. Analyses were done on all babies for whom a pulse oximetry reading was obtained. Sensitivity of pulse oximetry was 75·00% (95% CI 53·29-90·23) for critical cases and 49·06% (35·06-63·16) for all major congenital heart defects. In 35 cases, congenital heart defects were already suspected after antenatal ultrasonography, and exclusion of these reduced the sensitivity to 58·33% (27·67-84·83) for critical cases and 28·57% (14·64-46·30) for all cases of major congenital heart defects. False-positive results were noted for 169 (0·8%) babies (specificity 99·16%, 99·02-99·28), of which six cases were significant, but not major, congenital heart defects, and 40 were other illnesses that required urgent medical intervention.
Interpretation: Pulse oximetry is a safe, feasible test that adds value to existing screening. It identifies cases of critical congenital heart defects that go undetected with antenatal ultrasonography. The early detection of other diseases is an additional advantage.
Funding: National Institute for Health Research Health Technology Assessment programme.
Copyright © 2011 Elsevier Ltd. All rights reserved.
Comment in
-
Screening with pulse oximetry for congenital heart disease.Lancet. 2011 Aug 27;378(9793):749-50. doi: 10.1016/S0140-6736(11)61032-5. Epub 2011 Aug 4. Lancet. 2011. PMID: 21820731 No abstract available.
-
Congenital heart conditions: Clear benefits of pulse oximetry in neonates--one small step to improving postnatal cardiac care.Nat Rev Cardiol. 2011 Aug 23;8(10):539. doi: 10.1038/nrcardio.2011.131. Nat Rev Cardiol. 2011. PMID: 21862975 No abstract available.
-
Sensitivity of pulse oximetry for detection of critical congenital heart defects in newborn infants higher than that of antenatal ultrasound with few false positives.Evid Based Med. 2012 Apr;17(2):57-8. doi: 10.1136/ebmed-2011-100290. Epub 2011 Nov 28. Evid Based Med. 2012. PMID: 22127340 No abstract available.
-
Screening for congenital heart disease with newborn pulse oximetry.Lancet. 2012 Jan 28;379(9813):309-10; author reply 311. doi: 10.1016/S0140-6736(12)60135-4. Lancet. 2012. PMID: 22284648 No abstract available.
-
Screening for congenital heart disease with newborn pulse oximetry.Lancet. 2012 Jan 28;379(9813):310-1; author reply 311. doi: 10.1016/S0140-6736(12)60137-8. Lancet. 2012. PMID: 22284649 No abstract available.
-
Screening for congenital heart disease with newborn pulse oximetry.Lancet. 2012 Jan 28;379(9813):310; author reply 311. doi: 10.1016/S0140-6736(12)60136-6. Lancet. 2012. PMID: 22284650 No abstract available.
-
Pulse oximetry is beneficial in screening newborns for critical congenital heart disease.J Pediatr. 2012 Mar;160(3):529. doi: 10.1016/j.jpeds.2012.01.007. J Pediatr. 2012. PMID: 22329879 No abstract available.
-
Pulse oximetry screening for congenital heart disease.Ir Med J. 2012 Apr;105(4):122-3. Ir Med J. 2012. PMID: 22708231 No abstract available.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical