Raw data-based iterative reconstruction in body CTA: evaluation of radiation dose saving potential
Abstract
Objective: To evaluate prospectively, in patients undergoing body CTA, the radiation dose saving potential of raw data-based iterative reconstruction as compared to filtered back projection (FBP).
Methods: Twenty-five patients underwent thoraco-abdominal CTA with 128-slice dual-source CT, operating both tubes at 120 kV. Full-dose (FD) images were reconstructed with FBP and were compared to half-dose (HD) images with FBP and HD-images with sinogram-affirmed iterative reconstruction (SAFIRE), both reconstructed using data from only one tube-detector-system. Image quality and sharpness of the aortic contour were assessed. Vessel attenuation and noise were measured, contrast-to-noise-ratio was calculated.
Results: Noise as image quality deteriorating artefact occurred in 24/25 (96%) HD-FBP but not in FD-FBP and HD-raw data-based iterative reconstruction datasets (p < 0.001). Other artefacts occurred with similar prevalence among the datasets. Sharpness of the aortic contour was higher for FD-FBP and HD-raw data-based iterative reconstruction as compared to HD-FBP (p < 0.001). Aortoiliac attenuation was similar among all datasets (p > 0.05). Lowest noise was found for HD-raw data-based iterative reconstruction (7.23HU), being 9.4% lower than that in FD-FBP (7.98HU, p < 0.05) and 30.8% lower than in HD-FBP images (10.44HU, p < 0.001). Contrast-to-noise-ratio was lower in HD-FBP (p < 0.001) and higher in HD-raw data-based iterative reconstruction (p < 0.001) as compared to FD-FBP.
Conclusion: Intra-individual comparisons of image quality of body CTA suggest that raw data-based iterative reconstruction allows for dose reduction >50% while maintaining image quality. Key Points • Raw data-based iterative reconstruction reduces image noise and improves image quality as compared to filtered back projection • At a similar radiation dose, raw data-based iterative reconstruction improves the sharpness of vessel contours • In body CTA a dose reduction of >50% might be possible when using raw data-based iterative reconstructions, while image quality can be maintained.
Similar articles
-
Clinical evaluation and potential radiation dose reduction of the novel sinogram-affirmed iterative reconstruction technique (SAFIRE) in abdominal computed tomography angiography.Acad Radiol. 2013 Feb;20(2):165-72. doi: 10.1016/j.acra.2012.08.015. Epub 2012 Oct 23. Acad Radiol. 2013. PMID: 23099242
-
Intraindividual comparison of diagnostic performance in patients with hepatic metastasis of full-dose standard and half-dose iterative reconstructions with dual-source abdominal computed tomography.Invest Radiol. 2014 Apr;49(4):195-200. doi: 10.1097/RLI.0000000000000014. Invest Radiol. 2014. PMID: 24300843 Clinical Trial.
-
Accuracy of prospectively ECG-triggered very low-dose coronary dual-source CT angiography using iterative reconstruction for the detection of coronary artery stenosis: comparison with invasive catheterization.Eur Heart J Cardiovasc Imaging. 2014 Nov;15(11):1238-45. doi: 10.1093/ehjci/jeu113. Epub 2014 Jun 17. Eur Heart J Cardiovasc Imaging. 2014. PMID: 24939952
-
Iterative reconstruction for coronary CT angiography: finding its way.Int J Cardiovasc Imaging. 2012 Mar;28(3):613-20. doi: 10.1007/s10554-011-9832-3. Epub 2011 Feb 27. Int J Cardiovasc Imaging. 2012. PMID: 21359835 Review.
-
Deep Learning CT Image Reconstruction in Clinical Practice.Rofo. 2021 Mar;193(3):252-261. doi: 10.1055/a-1248-2556. Epub 2020 Dec 10. Rofo. 2021. PMID: 33302311 Review. English.
Cited by
-
Noise reduction in CT image using prior knowledge aware iterative denoising.Phys Med Biol. 2020 Nov 19;65(22):10.1088/1361-6560/abc231. doi: 10.1088/1361-6560/abc231. Phys Med Biol. 2020. PMID: 33065559 Free PMC article.
-
[SPECT/CT - Technical aspects and optimization possibilities].Radiologe. 2012 Jul;52(7):608-14. doi: 10.1007/s00117-011-2268-x. Radiologe. 2012. PMID: 22699407 German.
-
Image quality comparison of two adaptive statistical iterative reconstruction (ASiR, ASiR-V) algorithms and filtered back projection in routine liver CT.Br J Radiol. 2018 Jul;91(1088):20170655. doi: 10.1259/bjr.20170655. Epub 2018 Jun 6. Br J Radiol. 2018. PMID: 29848018 Free PMC article.
-
Vascular imaging: the evolving role of the multidisciplinary team meeting in peripheral vascular disease.Semin Intervent Radiol. 2014 Dec;31(4):320-9. doi: 10.1055/s-0034-1393968. Semin Intervent Radiol. 2014. PMID: 25435657 Free PMC article. Review.
-
Cohort study of Western Australia computed tomography utilisation patterns and their policy implications.BMC Health Serv Res. 2014 Nov 5;14:526. doi: 10.1186/s12913-014-0526-0. BMC Health Serv Res. 2014. PMID: 25370909 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous