Raw data-based iterative reconstruction in body CTA: evaluation of radiation dose saving potential

Affiliations
Free article

Raw data-based iterative reconstruction in body CTA: evaluation of radiation dose saving potential

Anna Winklehner et al. Eur Radiol. 2011 Dec.
Free article

Abstract

Objective: To evaluate prospectively, in patients undergoing body CTA, the radiation dose saving potential of raw data-based iterative reconstruction as compared to filtered back projection (FBP).

Methods: Twenty-five patients underwent thoraco-abdominal CTA with 128-slice dual-source CT, operating both tubes at 120 kV. Full-dose (FD) images were reconstructed with FBP and were compared to half-dose (HD) images with FBP and HD-images with sinogram-affirmed iterative reconstruction (SAFIRE), both reconstructed using data from only one tube-detector-system. Image quality and sharpness of the aortic contour were assessed. Vessel attenuation and noise were measured, contrast-to-noise-ratio was calculated.

Results: Noise as image quality deteriorating artefact occurred in 24/25 (96%) HD-FBP but not in FD-FBP and HD-raw data-based iterative reconstruction datasets (p < 0.001). Other artefacts occurred with similar prevalence among the datasets. Sharpness of the aortic contour was higher for FD-FBP and HD-raw data-based iterative reconstruction as compared to HD-FBP (p < 0.001). Aortoiliac attenuation was similar among all datasets (p > 0.05). Lowest noise was found for HD-raw data-based iterative reconstruction (7.23HU), being 9.4% lower than that in FD-FBP (7.98HU, p < 0.05) and 30.8% lower than in HD-FBP images (10.44HU, p < 0.001). Contrast-to-noise-ratio was lower in HD-FBP (p < 0.001) and higher in HD-raw data-based iterative reconstruction (p < 0.001) as compared to FD-FBP.

Conclusion: Intra-individual comparisons of image quality of body CTA suggest that raw data-based iterative reconstruction allows for dose reduction >50% while maintaining image quality. Key Points • Raw data-based iterative reconstruction reduces image noise and improves image quality as compared to filtered back projection • At a similar radiation dose, raw data-based iterative reconstruction improves the sharpness of vessel contours • In body CTA a dose reduction of >50% might be possible when using raw data-based iterative reconstructions, while image quality can be maintained.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Radiology. 2006 Dec;241(3):908-14 - PubMed
    1. Radiology. 2004 Mar;230(3):619-28 - PubMed
    1. AJR Am J Roentgenol. 2010 Sep;195(3):713-9 - PubMed
    1. Eur Radiol. 2011 Mar;21(3):636-43 - PubMed
    1. Eur Radiol. 2005 Feb;15(2):334-41 - PubMed

Publication types

MeSH terms