Technique for steering spinal cord stimulator electrode
- PMID: 21829146
- DOI: 10.1227/NEU.0b013e3182181eed
Technique for steering spinal cord stimulator electrode
Abstract
Background: Spinal cord stimulation is an established technology for management of chronic low back and leg pain when other surgical options have failed or are not feasible. Precise placement of the paddle-style electrode relative to the patient's distribution of pain can be difficult because of anatomic variation and the inherent limitations of electrode design.
Objective: To describe a surgical technique for adjustment of spinal cord stimulator epidural electrode location during placement or revision to optimize stimulation coverage.
Methods: We devised a method using a malleable disposable suture retrieval snare to precisely position an epidural electrode in either a midline or paramedian position during initial placement or revision. The snare is deployed and used to grasp a suture knot tied at the end of the electrode. The retriever is then used to position the electrode in the epidural space. Once satisfactory stimulation coverage is achieved, the snare is deployed to release the suture knot and the retriever is gently removed.
Results: We have used this method in 15 patients thus far to achieve optimal stimulation coverage. The total time required for electrode placement and repositioning, including the stimulation trial, has been substantially reduced. The patient-reported discomfort has also been significantly reduced. No adverse consequences of the technique have occurred.
Conclusion: We believe that this technique is a novel and effective method for accurately steering an epidural electrode for dorsal column stimulation for management of chronic pain.
Comment in
-
Technique for steering spinal cord stimulator electrode.Neurosurgery. 2012 May;70(5):E1336. doi: 10.1227/NEU.0b013e31824cacd4. Neurosurgery. 2012. PMID: 22286344 No abstract available.
Similar articles
-
Alternative approach to needle placement in spinal cord stimulator trial/implantation.Pain Physician. 2011 Jan-Feb;14(1):45-53. Pain Physician. 2011. PMID: 21267041
-
Spinal cord stimulation electrode design: prospective, randomized, controlled trial comparing percutaneous and laminectomy electrodes-part I: technical outcomes.Neurosurgery. 2002 Aug;51(2):381-9; discussion 389-90. Neurosurgery. 2002. PMID: 12182776 Clinical Trial.
-
The use of intraoperative electrophysiology for the placement of spinal cord stimulator paddle leads under general anesthesia.Neurosurgery. 2012 Jun;70(2 Suppl Operative):230-6. doi: 10.1227/NEU.0b013e318232ff29. Neurosurgery. 2012. PMID: 21869720 Review.
-
Alternative approach to needle placement in cervical spinal cord stimulator insertion.Pain Physician. 2011 Mar-Apr;14(2):195-210. Pain Physician. 2011. PMID: 21412374
-
[Treatment of chronic lumbago and radicular pain by spinal cord stimulation. Long-term results].Rev Rhum Ed Fr. 1994 Apr;61(4):271-7. Rev Rhum Ed Fr. 1994. PMID: 7920526 Review. French.
Cited by
-
Percutaneous spinal cord stimulation cylindrical lead placement for managing refractory neuropathic pain: A case series with an endoscopic-assisted approach.J Cent Nerv Syst Dis. 2024 Nov 29;16:11795735241302715. doi: 10.1177/11795735241302715. eCollection 2024. J Cent Nerv Syst Dis. 2024. PMID: 39629169 Free PMC article.
-
A review of organic and inorganic biomaterials for neural interfaces.Adv Mater. 2014 Mar 26;26(12):1846-85. doi: 10.1002/adma.201304496. Adv Mater. 2014. PMID: 24677434 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources