Major pitfalls in Doppler investigations are presented based on 340 evaluated cerebral Doppler examinations in infants. Substantial pitfalls may result from: A. Physics of sound waves and Doppler instruments (errors due to high pass filter cut off, aliasing, rapid image update). B. Quality and adjustment of the Doppler instrument (errors due to low sensitivity, inappropriate adjustment of Doppler controls, inadequate wall filter). C. Examination technique (errors due to an unfavourable angle of incidence or due to transducer-induced pressure: decrease predominantly in diastolic flow velocity-increase in maximum flow velocity in the straight sinus). D. Hemodynamics (errors due to spatial or temporal variations of the flow profile, pulsatility, non-uniform distribution of cerebral blood flow/CBF). E. Cerebral vascular anatomy (errors due to an unfavourable probe position as related to the three-dimensional arrangement of vessels, inadequate separation of closely adjacent vessels). F. Interpretation (flow velocity or Resistance Index/RI is taken to equal CBF, RI is taken to equal peripheral vascular resistance, one artery is taken to represent the cerebral circulation). Pitfalls may be avoided by using adequate means (low wall filter adjustment, high Doppler frequency, critical assessment of velocity spectra) to reduce the likelihood of errors occurring.
Similar articles
-
Major pitfalls in Doppler investigations. Part II. Low flow velocities and colour Doppler applications.Pediatr Radiol. 1990;20(5):304-10. doi: 10.1007/BF02013161. Pediatr Radiol. 1990. PMID: 2190150
-
Duplex-scanning of the deep venous drainage in the evaluation of blood flow velocity of the cerebral vascular system in infants.Pediatr Radiol. 1989;19(2):79-90. doi: 10.1007/BF02387891. Pediatr Radiol. 1989. PMID: 2646588
-
Flow velocity measurements as an index of cerebral blood flow. Validity of transcranial Doppler sonographic monitoring during cardiac surgery.Anesthesiology. 1994 Dec;81(6):1401-10. doi: 10.1097/00000542-199412000-00015. Anesthesiology. 1994. PMID: 7992909
-
The Doppler signal: where does it come from and what does it mean?AJR Am J Roentgenol. 1988 Sep;151(3):439-47. doi: 10.2214/ajr.151.3.439. AJR Am J Roentgenol. 1988. PMID: 2970215 Review.
-
Transcranial Doppler sonography. Part 1. Principles, technique, and normal appearances.Radiographics. 1995 Jan;15(1):179-91. doi: 10.1148/radiographics.15.1.7899596. Radiographics. 1995. PMID: 7899596 Review.
Cited by
-
Major pitfalls in Doppler investigations. Part II. Low flow velocities and colour Doppler applications.Pediatr Radiol. 1990;20(5):304-10. doi: 10.1007/BF02013161. Pediatr Radiol. 1990. PMID: 2190150
-
Regional cerebral blood flow velocity changes after indomethacin infusion in preterm infants.Arch Dis Child. 1992 Jul;67(7 Spec No):851-4. doi: 10.1136/adc.67.7_spec_no.851. Arch Dis Child. 1992. PMID: 1519988 Free PMC article.
-
Cerebrovascular responses to carbon dioxide in children anaesthetized with halothane and isoflurane.Can J Anaesth. 1991 Oct;38(7):817-25. doi: 10.1007/BF03036954. Can J Anaesth. 1991. PMID: 1742814 Clinical Trial.
-
The contribution of colour Doppler flow imaging to the study of cerebral haemodynamics in the neonate.Neuroradiology. 1993;35(4):300-6. doi: 10.1007/BF00602621. Neuroradiology. 1993. PMID: 8492901
-
Superior mesenteric artery and renal artery blood flow velocity measurements in neonates: technique and interobserver reliability.Pediatr Radiol. 1995;25(2):145-8. doi: 10.1007/BF02010331. Pediatr Radiol. 1995. PMID: 7596663